Characteristics of tissue temperature during ablation with THERMOCOOL SMARTTOUCH SF versus TactiCath versus QDOT MICRO catheters (Qmode and Qmode+): An in vivo porcine study

Naoto Otsuka iD, Yasuo Okumura, Sayaka Kuorkawa, Koichi Nagashima iD, Yuji Wakamatsu iD, Satoshi Hayashida iD, Kimie Ohkubo, Toshiko Nakai iD, Rie Takahashi, Yoshiki Taniguchi

Journal of Cardiovascular Electrophysiology · 2024

DOI: 10.1111/jce.16092 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Naoto Otsuka und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: Biosense Webster, Inc..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

High‐power short‐duration (HPSD) ablation at 50 W, guided by ablation index (AI) or lesion size index (LSI), and a 90 W/4 s very HSPD (vHPSD) setting are available for atrial fibrillation (AF) treatment. Yet, tissue temperatures during ablation with different catheters around venoatrial junction and collateral tissues remain unclear.

Methods

In this porcine study, we surgically implanted thermocouples on the epicardium near the superior vena cava (SVC), right pulmonary vein, and esophagus close to the inferior vena cava. We then compared tissue temperatures during 50W‐HPSD guided by AI 400 or LSI 5.0, and 90 W/4 s‐vHPSD ablation using THERMOCOOL SMARTTOUCH SF (STSF), TactiCath ablation catheter, sensor enabled (TacthCath), and QDOT MICRO (Qmode and Qmode+ settings) catheters.

Results

STSF produced the highest maximum tissue temperature (T max ), followed by TactiCath, and QDOT MICRO in Qmode and Qmode+ (62.7 ± 12.5°C, 58.0 ± 10.1°C, 50.0 ± 12.1°C, and 49.2 ± 8.4°C, respectively; p = .005), achieving effective transmural lesions. Time to lethal tissue temperature ≥50°C (t−T ≥ 50°C) was fastest in Qmode+, followed by TacthCath, STSF, and Qmode (4.3 ± 2.5, 6.4 ± 1.9, 7.1 ± 2.8, and 7.7 ± 3.1 s, respectively; p < .001). The catheter tip‐to‐thermocouple distance for lethal temperature (indicating lesion depth) from receiver operating characteristic curve analysis was deepest in STSF at 5.2 mm, followed by Qmode at 4.3 mm, Qmode+ at 3.1 mm, and TactiCath at 2.8 mm. Ablation at the SVC near the phrenic nerve led to sudden injury at t−T ≥ 50°C in all four settings. The esophageal adventitia injury was least deep with Qmode+ ablation (0.4 ± 0.1 vs. 0.8 ± 0.4 mm for Qmode, 0.9 ± 0.3 mm for TactiCath, and 1.1 ± 0.5 mm for STSF, respectively; p = .005), correlating with T max .

Conclusion

This study revealed distinct tissue temperature patterns during HSPD and vHPSD ablations with the three catheters, affecting lesion effectiveness and collateral damage based on T max and/or t−T ≥ 50°C. These findings provide key insights into the safety and efficacy of AF ablation with these four settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationPulmonary VeinsSwineAnimalsTemperatureVena Cava, SuperiorCathetersHot TemperatureTreatment Outcome

Stichwörter der Autor:innen

atrial fibrillationcatheter ablationesophageal injuryphrenic nerve injurytissue temperature

Autor:innen und Institutionen

  • Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Medical Research Support Center, Institute of Medical Science, Section of Laboratory for Animal Experiments, Nihon University School of Medicine, Tokyo, Japan.

ORCID-Profile: Naoto Otsuka, Koichi Nagashima, Yuji Wakamatsu, Satoshi Hayashida, Toshiko Nakai

Förderung

  • Biosense Webster, Inc.

Zitieren

APA

Otsuka, N., Okumura, Y., Kuorkawa, S., Nagashima, K., Wakamatsu, Y., Hayashida, S., Ohkubo, K., Nakai, T., Takahashi, R., & Taniguchi, Y. (2024). Characteristics of tissue temperature during ablation with THERMOCOOL SMARTTOUCH SF versus TactiCath versus QDOT MICRO catheters (Qmode and Qmode+): An in vivo porcine study. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16092

Vancouver

Otsuka N, Okumura Y, Kuorkawa S, Nagashima K, Wakamatsu Y, Hayashida S, et al. Characteristics of tissue temperature during ablation with THERMOCOOL SMARTTOUCH SF versus TactiCath versus QDOT MICRO catheters (Qmode and Qmode+): An in vivo porcine study. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16092

BibTeX
@article{Otsuka2024,
  author = {Otsuka, Naoto and Okumura, Yasuo and Kuorkawa, Sayaka and Nagashima, Koichi and Wakamatsu, Yuji and Hayashida, Satoshi and Ohkubo, Kimie and Nakai, Toshiko and Takahashi, Rie and Taniguchi, Yoshiki},
  title = {Characteristics of tissue temperature during ablation with THERMOCOOL SMARTTOUCH SF versus TactiCath versus QDOT MICRO catheters (Qmode and Qmode+): An in vivo porcine study},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2024},
  doi = {10.1111/jce.16092}
}
Kostenlos starten →