Validation of the Sysmex XN ‐ V hematology analyzer for feline specimens
Veterinary Clinical Pathology · 2024
DOI: 10.1111/vcp.13377 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Veterinary Clinical Pathology, 2024. Verfasst von Maud Guerlin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Département des Sciences Cliniques des Animaux de Compagnie et de Sport, Université de Toulouse, ENVT, Toulouse, France. Hauptthemen laut MEDLINE: Hematologic Tests. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The Sysmex XN‐V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT‐2000iV analyzer include an optic‐fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count.
Objective
We aimed to validate the Sysmex XN‐V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations.
Methods
Feline EDTA blood specimens were analyzed on the Sysmex XN‐V to evaluate repeatability, linearity, comparison with the XT‐2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT‐2000iV RIs.
Results
Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN‐V and XT‐2000iV analyzers was good (≥0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN‐V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT‐2000iV RIs can be used to interpret results from the Sysmex XN‐V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN‐V measurements after storage at 4 and 24°C were similar to those described for the Sysmex XT‐2000iV analyzer.
Conclusions
The performance of the Sysmex XN‐V analyzer was good and compared favorably with the Sysmex XT‐2000iV analyzer.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Département des Sciences Cliniques des Animaux de Compagnie et de Sport, Université de Toulouse, ENVT, Toulouse, France.
- CREFRE, Université de Toulouse, INSERM, ENVT, UPS, Toulouse, France.
- CRCT, Université de Toulouse, INSERM, UMR 1037, ENVT, Toulouse, France.
- Animal and Comparative Clinical Pathology, Toulouse, France.
ORCID-Profile: Nathalie Bourgès-Abella, Catherine Trumel
Zitieren
APA
Guerlin, M., Granat, F., Grebert, M., Braun, J. P., Geffré, A., Bourgès-Abella, N., & Trumel, C. (2024). Validation of the Sysmex XN ‐ V hematology analyzer for feline specimens. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.13377
Vancouver
Guerlin M, Granat F, Grebert M, Braun JP, Geffré A, Bourgès-Abella N, et al. Validation of the Sysmex XN ‐ V hematology analyzer for feline specimens. Veterinary Clinical Pathology. 2024. doi:10.1111/vcp.13377
BibTeX
@article{Guerlin2024,
author = {Guerlin, Maud and Granat, Fanny and Grebert, Margot and Braun, Jean-Pierre and Geffré, Anne and Bourgès-Abella, Nathalie and Trumel, Catherine},
title = {Validation of the Sysmex XN ‐ V hematology analyzer for feline specimens},
journal = {Veterinary Clinical Pathology},
year = {2024},
doi = {10.1111/vcp.13377}
}
EUVIMED