Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts
Artificial Organs · 2026-08 · 50 (8) · 1125-1135
DOI: 10.1111/aor.70152 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Artificial Organs, online erschienen am 01.05.2026. Verfasst von Sarah Linnemeier und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria. Hauptthemen laut MEDLINE: Heart-Assist Devices. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Österreichische Forschungsförderungsgesellschaft.
Abstract (Originalsprache)
Background
Hemocompatibility‐related adverse events still limit the broader adoption of rotodynamic blood pumps. This study aims to evaluate the potential of a multistage pump design to enhance hemocompatibility in left ventricular assist devices.
Methods
Computational fluid dynamic simulations were utilized to compare the pump performance and hemocompatibility of a Two‐Stage Pump to two equivalent single‐stage devices, designed for identical operating conditions. One single‐stage design features equal impeller and volute dimensions; the other, equal specific speed and specific diameter. Our simulation setup was validated against in vitro experiments of the pump performance of the Two‐Stage Pump.
Results
The Two‐Stage Pump achieved a substantial reduction in circumferential velocity (24% and 29% reduction). Although the two‐stage configuration thereby effectively reduces shear stress, it simultaneously translates to a longer residence time due to the larger priming volume. However, because the numerical blood damage estimation weighs shear stress exponentially more than residence time, the beneficial effect of reduced shear stress partially compensates for the longer residence time. As a result, the Two‐Stage Pump demonstrates a lower hemolysis index (12% and 14% reduction) compared to both single‐stage designs.
Conclusion
Our comparative study demonstrates that the two‐stage concept offers favorable hemolysis measures compared to single‐stage designs, as the reduction in peak shear stresses outweighs the disadvantage associated with prolonged residence time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
- Institute of Energy Systems and Thermodynamics, Vienna University of Technology, Vienna, Austria.
ORCID-Profile: Sarah Linnemeier
Förderung
- Österreichische Forschungsförderungsgesellschaft
Zitieren
APA
Linnemeier, S., Thamsen, B., Maric, S., Semlitsch, B., Torner, B., Zimpfer, D., & Granegger, M. (2026). Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts. Artificial Organs, 50(8), 1125-1135. https://doi.org/10.1111/aor.70152
Vancouver
Linnemeier S, Thamsen B, Maric S, Semlitsch B, Torner B, Zimpfer D, et al. Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts. Artificial Organs. 2026;50(8):1125-1135. doi:10.1111/aor.70152
BibTeX
@article{Linnemeier2026,
author = {Linnemeier, Sarah and Thamsen, Bente and Maric, Stefan and Semlitsch, Bernhard and Torner, Benjamin and Zimpfer, Daniel and Granegger, Marcus},
title = {Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts},
journal = {Artificial Organs},
year = {2026},
volume = {50},
number = {8},
pages = {1125--1135},
doi = {10.1111/aor.70152}
}
EUVIMED