EUVIMEDEUVIMED entdecken
Crossref

Multi‐Objective CFD Optimization of an Intermediate Diffuser Stage for PediaFlow Pediatric Ventricular Assist Device

Mansur Zhussupbekov iD, JingChun Wu, Greg W Burgreen, Scott Stelick, Jeongho Kim, James F Antaki iD

Artificial Organs · 2026

DOI: 10.1111/aor.70004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2026. Verfasst von Mansur Zhussupbekov und 5 weiteren Autor:innen. Beteiligte Einrichtung: Meinig School of Biomedical Engineering, Cornell University, Ithaca, New York, USA. Hauptthemen laut MEDLINE: Heart-Assist Devices, Prosthesis Design, Models, Cardiovascular. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: U.S. Army Medical Research Acquisition Activity, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Computational fluid dynamics (CFD) has become an essential design tool for ventricular assist devices (VADs), where the goal of maximizing performance often conflicts with biocompatibility. This tradeoff becomes even more pronounced in pediatric applications due to the stringent size constraints imposed by the smaller patient population. This study presents an automated CFD‐driven shape optimization of a new intermediate diffuser stage for the PediaFlow pediatric VAD, positioned immediately downstream of the impeller to improve pressure recovery.

Methods

We adopted a multi‐objective optimization approach to maximize pressure recovery while minimizing hemolysis. The proposed diffuser stage was isolated from the rest of the flow domain, enabling efficient evaluation of over 450 design variants using Sobol sequence, which yielded a Pareto front of nondominated solutions. The selected best candidate was further refined using a local T‐search algorithm. We then incorporated the optimized front diffuser into the full pump for CFD verification and in vitro validation.

Results

We identified critical dependencies where longer blades increased pressure recovery but also hemolysis, while the wrap angle showed a strong parabolic relationship with pressure recovery but a monotonic relationship with hemolysis. Counterintuitively, configurations with fewer blades (2, 3) consistently outperformed those with more blades (4, 5) in both metrics. The optimized two‐blade design enabled operation at lower pump speeds (14 000 vs. 16 000 RPM), improving hydraulic efficiency from 26.3% to 32.5% and reducing hemolysis by 31%.

Conclusion

This approach demonstrates that multi‐objective CFD optimization can systematically explore complex design spaces while balancing competing priorities of performance and hemocompatibility for pediatric VADs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesProsthesis DesignModels, CardiovascularHumansHydrodynamicsHemolysisChildComputer SimulationAlgorithms

Stichwörter der Autor:innen

diffuserefficiencyhemolysismulti‐objectiveoptimizationpediatricventricular assist device computational fluid dynamics

Autor:innen und Institutionen

  • Meinig School of Biomedical Engineering, Cornell University, Ithaca, New York, USA.
  • Advanced Design Optimization, LLC, Irvine, California, USA.
  • Center for Advanced Vehicular Systems, Mississippi State University, Starkville, Mississippi, USA.
  • Department of Biomedical Engineering, Daejeon Institute of Science and Technology, Daejeon, Korea.

ORCID-Profile: Mansur Zhussupbekov, James F Antaki

Förderung

  • U.S. Army Medical Research Acquisition Activity
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Zhussupbekov, M., Wu, J., Burgreen, G. W., Stelick, S., Kim, J., & Antaki, J. F. (2026). Multi‐Objective CFD Optimization of an Intermediate Diffuser Stage for PediaFlow Pediatric Ventricular Assist Device. Artificial Organs. https://doi.org/10.1111/aor.70004

Vancouver

Zhussupbekov M, Wu J, Burgreen GW, Stelick S, Kim J, Antaki JF. Multi‐Objective CFD Optimization of an Intermediate Diffuser Stage for PediaFlow Pediatric Ventricular Assist Device. Artificial Organs. 2026. doi:10.1111/aor.70004

BibTeX
@article{Zhussupbekov2026,
  author = {Zhussupbekov, Mansur and Wu, JingChun and Burgreen, Greg W and Stelick, Scott and Kim, Jeongho and Antaki, James F},
  title = {Multi‐Objective CFD Optimization of an Intermediate Diffuser Stage for PediaFlow Pediatric Ventricular Assist Device},
  journal = {Artificial Organs},
  year = {2026},
  doi = {10.1111/aor.70004}
}
Kostenlos starten →