In Vitro Hemolysis Evaluation of the Second Heart Assist Whisper Percutaneous Mechanical Circulatory Support Device

Chris H H Chan iD, Asma Farooqui iD, Preston Peak, Yaxin Wang, Sai Kode, Hadeel Al-Sahli, Zvonimir Krajcer, Sejal Chaudhari, Alex Richardson, Leslie Miller

Artificial Organs · 2025-07 · 49 (7) · 1119-1125

DOI: 10.1111/aor.14979 ↗

At a glance

Journal article in Artificial Organs, published online on 2025-02-24. Written by Chris H H Chan and 9 co-authors. Institution: Innovative Device and Engineering Applications Lab, The Texas Heart Institute, Houston, Texas, USA. Major MEDLINE topics: Hemolysis, Heart-Assist Devices. Cited 5 times so far (Crossref); 15 references. Funded by: Second Heart Assist Inc.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

High rotational pump speeds in percutaneous mechanical circulatory support (pMCS) devices frequently result in hemolysis. Consequently, in vitro hemolysis testing for newly developed devices is essential to mitigate this significant risk. This study compared the degree of hemolysis induced by the Second Heart Assist Whisper pMCS at its maximum pump speed, representing the worst‐case scenario, with that of the Impella CP, following American society for testing and materials standards.

Methods

Five Whisper devices were tested at 10000 RPM, while a comparator Impella CP was operated at 44 133 ± 606 RPM to match the Whisper's flow rate. Both devices were evaluated in two identical in vitro blood circulatory loops using citrated bovine blood. Hemolysis was analyzed by the tetramethylbenzidine method.

Results

The change in plasma free hemoglobin (Δ pf Hb) was significantly greater with the Impella CP than with the Whisper ( p < 0.01). Both devices caused a steady increase in pf Hb, with significant differences after 60 min of in vitro testing ( p < 0.01). The Whisper and Impella CP differed significantly in normalized index of hemolysis (0.088 ± 0.022 vs. 0.194 ± 0.029 g/100, p < 0.01) and modified index of hemolysis (9.72 ± 2.4 vs. 20.83 ± 3.5, p < 0.01).

Conclusion

The Whisper exhibited superior hemolytic performance to that of the Impella CP under identical hemodynamic conditions. Our newly designed blood circulatory loop and these benchmark values will aid future in vitro hemolysis testing for pMCS devices.

Subjects

MeSH terms (MEDLINE)

HemolysisHeart-Assist DevicesAnimalsCattleMaterials TestingProsthesis DesignHemoglobinsHumans

Author keywords

Impella CPhemolysisin vitro testingpercutaneous mechanical circulatory supportwhisper

Substances

Hemoglobins

Authors and institutions

  • Innovative Device and Engineering Applications Lab, The Texas Heart Institute, Houston, Texas, USA.
  • Interventional Cardiology, the Texas Heart Institute Center for Cardiovascular Care, The Texas Heart Institute, Houston, Texas, USA.
  • Second Heart Assist, Salt Lake City, Utah, USA.

ORCID profiles: Chris H H Chan, Asma Farooqui

Funding

  • Second Heart Assist Inc

Cite

APA

Chan, C. H. H., Farooqui, A., Peak, P., Wang, Y., Kode, S., Al-Sahli, H., Krajcer, Z., Chaudhari, S., Richardson, A., & Miller, L. (2025). In Vitro Hemolysis Evaluation of the Second Heart Assist Whisper Percutaneous Mechanical Circulatory Support Device. Artificial Organs, 49(7), 1119-1125. https://doi.org/10.1111/aor.14979

Vancouver

Chan CHH, Farooqui A, Peak P, Wang Y, Kode S, Al-Sahli H, et al. In Vitro Hemolysis Evaluation of the Second Heart Assist Whisper Percutaneous Mechanical Circulatory Support Device. Artificial Organs. 2025;49(7):1119-1125. doi:10.1111/aor.14979

BibTeX
@article{Chan2025,
  author = {Chan, Chris H H and Farooqui, Asma and Peak, Preston and Wang, Yaxin and Kode, Sai and Al-Sahli, Hadeel and Krajcer, Zvonimir and Chaudhari, Sejal and Richardson, Alex and Miller, Leslie},
  title = {In Vitro Hemolysis Evaluation of the Second Heart Assist Whisper Percutaneous Mechanical Circulatory Support Device},
  journal = {Artificial Organs},
  year = {2025},
  volume = {49},
  number = {7},
  pages = {1119--1125},
  doi = {10.1111/aor.14979}
}
Start for free →