A parametric study on pulse duplicator design and valve hemodynamics
Artificial Organs · 2024-09 · 48 (9) · 977-987
DOI: 10.1111/aor.14757 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 23.04.2024. Verfasst von Caroline C Smid und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Composite Materials and Adaptive Structures, ETH Zurich, Zürich, Switzerland. Hauptthemen laut MEDLINE: Hemodynamics, Heart Valve Prosthesis, Prosthesis Design, Aortic Valve. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Zurich Heart, Hochschulmedizin Zürich, Switzerland, ETH Zürich, Switzerland, ETHeart, Open ETH, ETH Board, Switzerland.
Abstract (Originalsprache)
Background
In vitro assessment is mandatory for artificial heart valve development. This study aims to investigate the effects of pulse duplicator features on valve responsiveness, conduct a sensitivity analysis across valve prosthesis types, and contribute on the development of versatile pulse duplicator systems able to perform reliable prosthetic aortic valve assessment under physiologic hemodynamic conditions.
Methods
A reference pulse duplicator was established based on literature. Further optimization process led to new designs that underwent a parametric study, also involving different aortic valve prostheses. These designs were evaluated on criteria such as mean pressure differential and pulse pressure (assessed from high‐fidelity pressure measurements), valve opening and closing behavior, flow, and regurgitation. Finally, the resulting optimized setup was tested under five different hemodynamic settings simulating a range of physiologic and pathologic conditions.
Results
The results show that both, pulse duplicator design and valve type significantly influence aortic and ventricular pressure, flow, and valve kinematic response. The optimal design comprised key features such as a compliance chamber and restrictor for diastolic pressure maintenance and narrow pulse pressure. Additionally, an atrial reservoir was included to prevent atrial–aortic interference, and a bioprosthetic valve was used in mitral position to avoid delayed valve closing effects.
Conclusion
This study showed that individual pulse duplicator features can have a significant effect on valve's responsiveness. The optimized versatile pulse duplicator replicated physiologic and pathologic aortic valve hemodynamic conditions, serving as a reliable characterization tool for assessing and optimizing aortic valve performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratory of Composite Materials and Adaptive Structures, ETH Zurich, Zürich, Switzerland.
- Translational Cardiovascular Technologies, ETH Zurich, Zürich, Switzerland.
- Department for Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum der Charité, Charité Universitätsmedizin Berlin, Berlin, Germany.
- German Center for Cardiovascular Research (DZHK), Partner Site Berlin, Berlin, Germany.
ORCID-Profile: Caroline C Smid, Georgios A Pappas, Volkmar Falk, Paolo Ermanni, Nikola Cesarovic
Förderung
- Zurich Heart, Hochschulmedizin Zürich, Switzerland
- ETH Zürich, Switzerland
- ETHeart, Open ETH, ETH Board, Switzerland
- Eidgenössische Technische Hochschule Zürich
Zitieren
APA
Smid, C. C., Pappas, G. A., Falk, V., Ermanni, P., & Cesarovic, N. (2024). A parametric study on pulse duplicator design and valve hemodynamics. Artificial Organs, 48(9), 977-987. https://doi.org/10.1111/aor.14757
Vancouver
Smid CC, Pappas GA, Falk V, Ermanni P, Cesarovic N. A parametric study on pulse duplicator design and valve hemodynamics. Artificial Organs. 2024;48(9):977-987. doi:10.1111/aor.14757
BibTeX
@article{Smid2024,
author = {Smid, Caroline C and Pappas, Georgios A and Falk, Volkmar and Ermanni, Paolo and Cesarovic, Nikola},
title = {A parametric study on pulse duplicator design and valve hemodynamics},
journal = {Artificial Organs},
year = {2024},
volume = {48},
number = {9},
pages = {977--987},
doi = {10.1111/aor.14757}
}
EUVIMED