A parametric study on pulse duplicator design and valve hemodynamics

Caroline C Smid iD, Georgios A Pappas iD, Volkmar Falk iD, Paolo Ermanni iD, Nikola Cesarovic iD

Artificial Organs · 2024-09 · 48 (9) · 977-987

DOI: 10.1111/aor.14757 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 23.04.2024. Verfasst von Caroline C Smid und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Composite Materials and Adaptive Structures, ETH Zurich, Zürich, Switzerland. Hauptthemen laut MEDLINE: Hemodynamics, Heart Valve Prosthesis, Prosthesis Design, Aortic Valve. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Zurich Heart, Hochschulmedizin Zürich, Switzerland, ETH Zürich, Switzerland, ETHeart, Open ETH, ETH Board, Switzerland.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In vitro assessment is mandatory for artificial heart valve development. This study aims to investigate the effects of pulse duplicator features on valve responsiveness, conduct a sensitivity analysis across valve prosthesis types, and contribute on the development of versatile pulse duplicator systems able to perform reliable prosthetic aortic valve assessment under physiologic hemodynamic conditions.

Methods

A reference pulse duplicator was established based on literature. Further optimization process led to new designs that underwent a parametric study, also involving different aortic valve prostheses. These designs were evaluated on criteria such as mean pressure differential and pulse pressure (assessed from high‐fidelity pressure measurements), valve opening and closing behavior, flow, and regurgitation. Finally, the resulting optimized setup was tested under five different hemodynamic settings simulating a range of physiologic and pathologic conditions.

Results

The results show that both, pulse duplicator design and valve type significantly influence aortic and ventricular pressure, flow, and valve kinematic response. The optimal design comprised key features such as a compliance chamber and restrictor for diastolic pressure maintenance and narrow pulse pressure. Additionally, an atrial reservoir was included to prevent atrial–aortic interference, and a bioprosthetic valve was used in mitral position to avoid delayed valve closing effects.

Conclusion

This study showed that individual pulse duplicator features can have a significant effect on valve's responsiveness. The optimized versatile pulse duplicator replicated physiologic and pathologic aortic valve hemodynamic conditions, serving as a reliable characterization tool for assessing and optimizing aortic valve performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemodynamicsHeart Valve ProsthesisProsthesis DesignAortic ValveHumansModels, CardiovascularBioprosthesis

Stichwörter der Autor:innen

aortic valvebioprosthetic valvehemodynamic settingsmechanical valvepolymeric valvepulse duplicatorvalve characterization

Autor:innen und Institutionen

  • Laboratory of Composite Materials and Adaptive Structures, ETH Zurich, Zürich, Switzerland.
  • Translational Cardiovascular Technologies, ETH Zurich, Zürich, Switzerland.
  • Department for Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum der Charité, Charité Universitätsmedizin Berlin, Berlin, Germany.
  • German Center for Cardiovascular Research (DZHK), Partner Site Berlin, Berlin, Germany.

ORCID-Profile: Caroline C Smid, Georgios A Pappas, Volkmar Falk, Paolo Ermanni, Nikola Cesarovic

Förderung

  • Zurich Heart, Hochschulmedizin Zürich, Switzerland
  • ETH Zürich, Switzerland
  • ETHeart, Open ETH, ETH Board, Switzerland
  • Eidgenössische Technische Hochschule Zürich

Zitieren

APA

Smid, C. C., Pappas, G. A., Falk, V., Ermanni, P., & Cesarovic, N. (2024). A parametric study on pulse duplicator design and valve hemodynamics. Artificial Organs, 48(9), 977-987. https://doi.org/10.1111/aor.14757

Vancouver

Smid CC, Pappas GA, Falk V, Ermanni P, Cesarovic N. A parametric study on pulse duplicator design and valve hemodynamics. Artificial Organs. 2024;48(9):977-987. doi:10.1111/aor.14757

BibTeX
@article{Smid2024,
  author = {Smid, Caroline C and Pappas, Georgios A and Falk, Volkmar and Ermanni, Paolo and Cesarovic, Nikola},
  title = {A parametric study on pulse duplicator design and valve hemodynamics},
  journal = {Artificial Organs},
  year = {2024},
  volume = {48},
  number = {9},
  pages = {977--987},
  doi = {10.1111/aor.14757}
}
Kostenlos starten →