Transient numerical simulation of hemodynamics in bioprosthetic heart valves: insights into valve sizing and thrombosis risk
International Journal of Computer Assisted Radiology and Surgery · 2026
DOI: 10.1007/s11548-026-03577-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2026. Verfasst von M Berger und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Technologies, MCI - The Entrepreneurial School, Innsbruck, Austria. manuel.berger@mci.edu. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Bioprosthesis, Hemodynamics, Computer Simulation. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: MCI Management Center Innsbruck – Internationale Hochschule GmbH.
Abstract (Originalsprache)
Purpose
This study aims to evaluate the hemodynamic effect of tilting after implantation of a bioprosthetic heart valve in an enlarged aortic annulus using computational fluid dynamics (CFD). The objective is to define the cost of excessive enlargement in transprosthetical flow and identify the optimal prosthesis size and implantation angle to reduce thrombosis and early degeneration.
Methods
Virtual implantation of a 23-mm bioprosthetic model was conducted in a prespecified virtual aortic annulus of 23 mm. CFD simulations were conducted to analyze the hemodynamic parameters, including wall shear stress and turbulent kinetic energy, for bioprosthetic heart valves of the 23 mm (perpendicular implantation), 25 mm, and 27 mm (tilted implantation at different angles) after virtual annular enlargement. The simulations utilized transient flow models and mesh convergence studies, to ensure numerical accuracy and clinical relevance.
Results
The 23-mm valve implanted without annular enlargement and in an aligned fashion to the annulus exhibited higher wall shear stress (WSS) and shear stress (SS) values than the 25-mm valve implanted at 12° tilted position and better hemodynamics than the 27-mm valve at 25° tilted position. The 25-mm valve after annular enlargement and implantation at 12° tilted position achieved the best hemodynamic performance together with a 23% and 7% reduction of WSS and SS as compared to the 23-mm valve.
Conclusions
Slight oversizing and tilting after annular enlargement yields the best performance of an aortic bioprosthetic valve with lowest WSS and balanced turbulence.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Technologies, MCI - The Entrepreneurial School, Innsbruck, Austria. manuel.berger@mci.edu.
- Department of Medical Technologies, MCI - The Entrepreneurial School, Innsbruck, Austria.
- Department of Industrial Engineering & Management, MCI - The Entrepreneurial School, Innsbruck, Austria.
- Department of Cardiac Surgery, Medical University of Innsbruck, Innsbruck, Austria.
Förderung
- MCI Management Center Innsbruck – Internationale Hochschule GmbH
Zitieren
APA
Berger, M., Golks, J., Gleissner, L., Senfter, T., Mayerl, C., Bonaros, N., Tepeköylü, C. G., Stastny, L., Grimm, M., & Pillei, M. (2026). Transient numerical simulation of hemodynamics in bioprosthetic heart valves: insights into valve sizing and thrombosis risk. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-026-03577-4
Vancouver
Berger M, Golks J, Gleissner L, Senfter T, Mayerl C, Bonaros N, et al. Transient numerical simulation of hemodynamics in bioprosthetic heart valves: insights into valve sizing and thrombosis risk. International Journal of Computer Assisted Radiology and Surgery. 2026. doi:10.1007/s11548-026-03577-4
BibTeX
@article{Berger2026,
author = {Berger, M and Golks, J and Gleissner, L and Senfter, T and Mayerl, Ch and Bonaros, N and Tepeköylü, C G and Stastny, L and Grimm, M and Pillei, M},
title = {Transient numerical simulation of hemodynamics in bioprosthetic heart valves: insights into valve sizing and thrombosis risk},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2026},
doi = {10.1007/s11548-026-03577-4}
}
EUVIMED