A Time-Walk and Timing-Shift Correction Method for Dual-Ended Readout TOF-DOI PET Detectors
IEEE Transactions on Radiation and Plasma Medical Sciences · 2025-03 · 9 (3) · 277-283
DOI: 10.1109/trpms.2024.3497774 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2025-03. Verfasst von Haibo Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of California at Davis, Davis, CA 95616 USA. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Positron emission tomography (PET) detectors suffer from time-walk when the leading edge discriminator is employed for timing pick-off as well as a timing-shift when thick crystals are utilized due to the depth-of-interaction (DOI) effect. In this study, a combined time-walk and timing-shift correction method was proposed for dual-ended readout PET detectors. To evaluate the proposed method, a pair of dual-ended readout PET detectors was constructed. Each detector was based on two Hamamatsu S14161-3050-08 silicon photomultiplier (SiPM) arrays coupled to both ends of an 8 × 8 arrays of 3 . 1 × 3 . 1 × 20 m m 3 LYSO crystals with a 3.2 mm pitch. By employing the relationship between the energies and detected time differences of events, the time-walk and timing-shift were effectively corrected. The coincidence time resolution of the two detectors improved from 260 . 7 ± 1 . 0 ps to 229 . 4 ± 1 . 0 ps when a 400-650 keV energy window was used to select events. These results demonstrate the effectiveness of the proposed time-walk and timing-shift correction method.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of California at Davis, Davis, CA 95616 USA.
Förderung
- NIBIB NIH HHS
- National Institutes of Health
Zitieren
APA
Wang, H., Xie, J., Qi, J., Cherry, S. R., & Du, J. (2025). A Time-Walk and Timing-Shift Correction Method for Dual-Ended Readout TOF-DOI PET Detectors. IEEE Transactions on Radiation and Plasma Medical Sciences, 9(3), 277-283. https://doi.org/10.1109/trpms.2024.3497774
Vancouver
Wang H, Xie J, Qi J, Cherry SR, Du J. A Time-Walk and Timing-Shift Correction Method for Dual-Ended Readout TOF-DOI PET Detectors. IEEE Transactions on Radiation and Plasma Medical Sciences. 2025;9(3):277-283. doi:10.1109/trpms.2024.3497774
BibTeX
@article{Wang2025,
author = {Wang, Haibo and Xie, Jiahao and Qi, Jinyi and Cherry, Simon R and Du, Junwei},
title = {A Time-Walk and Timing-Shift Correction Method for Dual-Ended Readout TOF-DOI PET Detectors},
journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
year = {2025},
volume = {9},
number = {3},
pages = {277--283},
doi = {10.1109/trpms.2024.3497774}
}
EUVIMED