EUVIMEDEUVIMED entdecken
Crossref

Water-Fat Separation for the Knee on a 50 mT Portable MRI Scanner

Cai Wan, Wei He, Zheng Xu

IEEE Transactions on Biomedical Engineering · 2024

DOI: 10.1109/tbme.2023.3347441 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Biomedical Engineering, 2024. Verfasst von Cai Wan und 2 weiteren Autor:innen. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Adipose Tissue, Phantoms, Imaging. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 58 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Technological Innovation and Application Development in Chongqing, Chongqing Science and Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: The Dixon method is frequently employed in clinical and scientific research for fat suppression, because it has lower sensitivity to static magnetic field inhomogeneity compared to chemical shift selective saturation or its variants and maintains image signal-to-noise ratio (SNR). Recently, research on very-low-field (VLF < 100 mT) magnetic resonance imaging (MRI) has regained popularity. However, there is limited literature on water-fat separation in VLF MRI. Here, we present a modified two-point Dixon method specifically designed for VLF MRI. METHODS: Most experiments were performed on a homemade 50 mT portable MRI scanner. The receiving coil adopted a homemade quadrature receiving coil. The data were acquired using spin-echo and gradient-echo sequences. We considered the T2* effect, and added priori information to existing two-point Dixon method. Then, the method used regional iterative phasor extraction (RIPE) to extract the error phasor. Finally, least squares solutions for water and fat were obtained and fat signal fraction was calculated. RESULTS: For phantom evaluation, water-only and fat-only images were obtained and the local fat signal fractions were calculated, with two samples being 0.94 and 0.93, respectively. For knee imaging, cartilage, muscle and fat could be clearly distinguished. The water-only images were able to highlight areas such as cartilage that could not be easily distinguished without separation. CONCLUSION: This work has demonstrated the feasibility of using a 50 mT MRI scanner for water-fat separation. SIGNIFICANCE: To the best of our knowledge, this is the first reported result of water-fat separation at a 50 mT portable MRI scanner.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingAdipose TissuePhantoms, ImagingHumansBody WaterKneeKnee JointImage Processing, Computer-AssistedEquipment Design

Förderung

  • National Natural Science Foundation of China
  • Technological Innovation and Application Development in Chongqing
  • Chongqing Science and Health
  • Shenzhen Science and Technology Innovation Commission

Zitieren

APA

Wan, C., He, W., & Xu, Z. (2024). Water-Fat Separation for the Knee on a 50 mT Portable MRI Scanner. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/tbme.2023.3347441

Vancouver

Wan C, He W, Xu Z. Water-Fat Separation for the Knee on a 50 mT Portable MRI Scanner. IEEE Transactions on Biomedical Engineering. 2024. doi:10.1109/tbme.2023.3347441

BibTeX
@article{Wan2024,
  author = {Wan, Cai and He, Wei and Xu, Zheng},
  title = {Water-Fat Separation for the Knee on a 50 mT Portable MRI Scanner},
  journal = {IEEE Transactions on Biomedical Engineering},
  year = {2024},
  doi = {10.1109/tbme.2023.3347441}
}
Kostenlos starten →