A Differential Impedance Flow Cytometry Front-End With Baseline Current Cancellation

Siyuan Yu, Louis Marun, Matthew L Johnston

IEEE Transactions on Biomedical Circuits and Systems · 2026

DOI: 10.1109/tbcas.2025.3585089 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Biomedical Circuits and Systems, 2026. Verfasst von Siyuan Yu und 2 weiteren Autor:innen. Hauptthemen laut MEDLINE: Flow Cytometry. Die Arbeit nennt 30 Literaturangaben. Gefördert von: Center for Design of Analog-Digital Integrated Circuits.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this work, we present a high-performance analog front-end (AFE) circuit for impedance-based flow cytometry readout. The AFE is designed to interface to a three-electrode sensor topology using center electrode excitation and differential current output. To satisfy the needs of a differential high gain signal path, we propose a digitally tunable and calibrated cancellation current generation path to remove the baseline current injected into the transimpedance amplifier (TIA) stages. This prevents TIA saturation and allows for higher gain. Consequently, the AFE is more power efficient while maintaining better noise and interference rejection. The proposed circuit is designed and fabricated in a 180 nm CMOS process. It covers an excitation frequency range of 0.5 MHz to 10 MHz and consumes 15.6 mW during nominal operation. Digital calibration is implemented using an off-chip ADC and automated calibration algorithm. Measurement results show that at 1 MHz excitation, the AFE achieves 1.7 pA/$\sqrt{\text{Hz}}$ input-referred current noise density with floating inputs. The AFE achieves detection of 3 um diameter particles in a microfluidic flow cell, demonstrating its performance and practicality for impedance flow cytometry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Flow CytometryElectric ImpedanceEquipment DesignAlgorithmsSignal Processing, Computer-AssistedElectrodesCalibration

Förderung

  • Center for Design of Analog-Digital Integrated Circuits

Zitieren

APA

Yu, S., Marun, L., & Johnston, M. L. (2026). A Differential Impedance Flow Cytometry Front-End With Baseline Current Cancellation. IEEE Transactions on Biomedical Circuits and Systems. https://doi.org/10.1109/tbcas.2025.3585089

Vancouver

Yu S, Marun L, Johnston ML. A Differential Impedance Flow Cytometry Front-End With Baseline Current Cancellation. IEEE Transactions on Biomedical Circuits and Systems. 2026. doi:10.1109/tbcas.2025.3585089

BibTeX
@article{Yu2026,
  author = {Yu, Siyuan and Marun, Louis and Johnston, Matthew L},
  title = {A Differential Impedance Flow Cytometry Front-End With Baseline Current Cancellation},
  journal = {IEEE Transactions on Biomedical Circuits and Systems},
  year = {2026},
  doi = {10.1109/tbcas.2025.3585089}
}
Kostenlos starten →