Organ dose modulation: a useful tool for dose reduction in clinical practice?

Nienke Wentink, Margo van Gent, Chris Borns iD

British Journal of Radiology · 01.07.2026 · 99 (1183) · 1405-1413

DOI: 10.1093/bjr/tqag098 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, online erschienen am 04.05.2026. Verfasst von Nienke Wentink und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Frisius Medical Center, Leeuwarden 8901BR, The Netherlands. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed. Die Arbeit nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To quantify the reduction in effective dose achieved through organ dose modulation (ODM) on a General Electric Revolution Apex CT scanner, while preserving image quality.

Methods

First, the noise index (NI) corresponding to comparable image quality—defined by signal-to-noise ratio (SNR)—was determined using a PMMA cylindrical phantom (CTDI phantom), both with and without ODM. Scan protocols for the thorax, abdomen, and head were evaluated. Next, radiation dose was measured at every 10° scan angle using a 10 cm ionization chamber for each protocol, with and without ODM. Finally, effective dose was estimated using the angle-dependent dose measurements in Monte Carlo simulation software (PCXMC), for tube voltages of 80, 100, and 120 kVp.

Results

Effective dose was reduced by 12-14 ± 1% for thorax, 12-13 ± 1% for abdomen, when using ODM. No significant change in effective dose was found for CT head. Notably, radiosensitive organs such as the breasts and eye lenses received 19-21 ± 1% and 25 ± 7% (upper-bound estimate) less dose, respectively.

Conclusions

This study demonstrates that ODM could significantly reduce both effective and organ-specific radiation doses without compromising image quality, especially when applied to the whole scan field. Advances in knowledge This study demonstrates with dose measurements with 10° angular accuracy that for this specific implementation of ODM, it should be applied to the whole scan field for both dose reduction and consistent image quality. This finding is contradictory to the vendor’s recommendation to apply ODM only to an area with a specific radiosensitive organ, such as the breasts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageTomography, X-Ray ComputedHumansPhantoms, ImagingMonte Carlo MethodSignal-To-Noise RatioHeadTomography Scanners, X-Ray Computed

Stichwörter der Autor:innen

CTimage qualityorgan based tube current modulation (OBTCM)organ dose modulation (ODM)protocol optimisationradiation dose reduction

Autor:innen und Institutionen

  • Department of Medical Physics, Frisius Medical Center, Leeuwarden 8901BR, The Netherlands.
  • Department of Medical Imaging and Radiation Therapy, Hanze University of Applied Sciences, Groningen 9714 CA, The Netherlands.

ORCID-Profile: Chris Borns

Zitieren

APA

Wentink, N., van Gent, M., & Borns, C. (2026). Organ dose modulation: a useful tool for dose reduction in clinical practice?. British Journal of Radiology, 99(1183), 1405-1413. https://doi.org/10.1093/bjr/tqag098

Vancouver

Wentink N, van Gent M, Borns C. Organ dose modulation: a useful tool for dose reduction in clinical practice?. British Journal of Radiology. 2026;99(1183):1405-1413. doi:10.1093/bjr/tqag098

BibTeX
@article{Wentink2026,
  author = {Wentink, Nienke and van Gent, Margo and Borns, Chris},
  title = {Organ dose modulation: a useful tool for dose reduction in clinical practice?},
  journal = {British Journal of Radiology},
  year = {2026},
  volume = {99},
  number = {1183},
  pages = {1405--1413},
  doi = {10.1093/bjr/tqag098}
}
Kostenlos starten →