The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter

Ioannis A Tsalafoutas iD, Shady AlKhazzam, Mohammed Hassan Kharita

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14356 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Ioannis A Tsalafoutas und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Section, Occupational Health and Safety Department, Hamad Medical Corporation, Doha, Qatar. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiation Dosage, Tomography, X-Ray Computed, Tomography Scanners, X-Ray Computed. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose. Material & Methods A dedicated phantom (Mercury 4.0) was scanned using two of the most frequently used clinical scanning protocols (chest and abdomen‐pelvis). The preset protocol settings were used as starting points (reference conditions). Scan direction, scan mode (helical vs. axial), total beam width, tube potential (kVp), and ATCM settings were then modified individually to understand their impact on radiation dose and image quality. Regarding the ATCM settings, the SmartmA minimum and maximum mA limits, and the noise index (NI) values were varied. As surrogates of patient dose, the CTDI vol and DLP values of each scan were used. As surrogates of image quality were used the image noise and the detectability index ( d’ ) of five different materials (air, solid water, polystyrene, iodine, and bone) embedded in the Mercury phantom calculated with the ImQuest software.

Results

The scanning direction did not have any effect on ATCM curves, unlike what has been observed in CT scanners from other manufacturers. Total beam width does matter, however, the SmartmA limit settings and kVp selection had the greatest impact on image quality and dose. It was seen that improper minimum mA limit settings practically invalidated the ATCM operation. In contrast, when full modulation was allowed without restrictions, noise standard deviation, and detectability index became much more consistent across the wide range of phantom diameters. For lower kVp settings an impressive dose reduction was observed that requires further investigation.

Conclusion

SmartmA is a tool that if not properly used may increase the patient doses considerably. Therefore, its settings should be carefully adjusted for each preset different clinical protocol.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiation DosageTomography, X-Ray ComputedTomography Scanners, X-Ray ComputedImage Processing, Computer-AssistedHumansRadiographic Image Interpretation, Computer-AssistedSignal-To-Noise Ratio

Stichwörter der Autor:innen

CTautomatic tube current modulationdetectability indeximage noisemercury phantomquality control

Autor:innen und Institutionen

  • Medical Physics Section, Occupational Health and Safety Department, Hamad Medical Corporation, Doha, Qatar.

ORCID-Profile: Ioannis A Tsalafoutas

Zitieren

APA

Tsalafoutas, I. A., AlKhazzam, S., & Kharita, M. H. (2024). The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14356

Vancouver

Tsalafoutas IA, AlKhazzam S, Kharita MH. The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14356

BibTeX
@article{Tsalafoutas2024,
  author = {Tsalafoutas, Ioannis A and AlKhazzam, Shady and Kharita, Mohammed Hassan},
  title = {The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14356}
}
Kostenlos starten →