Calculation method for novel upright CT scanner isodose curves
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14377 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Michael W Kissick und 6 weiteren Autor:innen. Beteiligte Einrichtung: Leo Cancer Care, Ltd., Middleton, Wisconsin, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Tomography, X-Ray Computed, Tomography Scanners, X-Ray Computed. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
A computational method based on Monte‐Carlo calculations is presented and used to calculate isodose curves for a new upright and tilting CT scanner useful for radiation protection purposes.
Methods
The TOPAS code platform with imported CAD files for key components was used to construct a calculation space for the scanner. A sphere of water acts as the patient would by creating scatter out of the bore. Maximum intensity dose maps are calculated for various possible tilt angles to make sure radiation protection for site planning uses the maximum possible dose everywhere.
Results
The resulting maximum intensity isodose lines are more rounded than ones for just a single tilt angle and so closer to isotropic. These maximum intensity curves are closer to the isotropic assumption used in CTDI or DLP based methods of site planning and radiation protection. The isodose lines are similar to those of a standard CT scanner, just tilted upwards. There is more metal above the beam that lessens the dose above versus below isocenter.
Conclusion
Aside from the orientation, this upright scanner is very similar to a typical CT scanner, and nothing different for shielding needs to be done for this new upright tilting CT scanner, because an isotropic scatter source is often assumed for any CT scanner.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Leo Cancer Care, Ltd., Middleton, Wisconsin, USA.
- Department of Medical Physics, University of Wisconsin - Madison, Madison, Wisconsin, USA.
- Engineering and Physical Sciences, University of Surrey, Guildford, UK.
- Medical Radiation Physics, National Physical Laboratory, Teddington, UK.
ORCID-Profile: Andries N Schreuder
Zitieren
APA
Kissick, M. W., Panaino, C., Criscuolo, A., Hayes, J., Hoffman, C., Mackie, R., & Schreuder, A. N. (2024). Calculation method for novel upright CT scanner isodose curves. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14377
Vancouver
Kissick MW, Panaino C, Criscuolo A, Hayes J, Hoffman C, Mackie R, et al. Calculation method for novel upright CT scanner isodose curves. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14377
BibTeX
@article{Kissick2024,
author = {Kissick, Michael W and Panaino, Costanza and Criscuolo, Anthony and Hayes, John and Hoffman, Carson and Mackie, Rockwell and Schreuder, Andries N},
title = {Calculation method for novel upright CT scanner isodose curves},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14377}
}
EUVIMED