Calculation method for novel upright CT scanner isodose curves

Michael W Kissick, Costanza Panaino, Anthony Criscuolo, John Hayes, Carson Hoffman, Rockwell Mackie, Andries N Schreuder iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14377 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Michael W Kissick und 6 weiteren Autor:innen. Beteiligte Einrichtung: Leo Cancer Care, Ltd., Middleton, Wisconsin, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Tomography, X-Ray Computed, Tomography Scanners, X-Ray Computed. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

A computational method based on Monte‐Carlo calculations is presented and used to calculate isodose curves for a new upright and tilting CT scanner useful for radiation protection purposes.

Methods

The TOPAS code platform with imported CAD files for key components was used to construct a calculation space for the scanner. A sphere of water acts as the patient would by creating scatter out of the bore. Maximum intensity dose maps are calculated for various possible tilt angles to make sure radiation protection for site planning uses the maximum possible dose everywhere.

Results

The resulting maximum intensity isodose lines are more rounded than ones for just a single tilt angle and so closer to isotropic. These maximum intensity curves are closer to the isotropic assumption used in CTDI or DLP based methods of site planning and radiation protection. The isodose lines are similar to those of a standard CT scanner, just tilted upwards. There is more metal above the beam that lessens the dose above versus below isocenter.

Conclusion

Aside from the orientation, this upright scanner is very similar to a typical CT scanner, and nothing different for shielding needs to be done for this new upright tilting CT scanner, because an isotropic scatter source is often assumed for any CT scanner.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodTomography, X-Ray ComputedTomography Scanners, X-Ray ComputedHumansRadiation ProtectionPhantoms, ImagingRadiation DosageAlgorithmsRadiotherapy Planning, Computer-AssistedImage Processing, Computer-Assisted

Stichwörter der Autor:innen

CT isodoseCT upright tilting gantryTOPAS Monte‐Carlo modeling

Autor:innen und Institutionen

  • Leo Cancer Care, Ltd., Middleton, Wisconsin, USA.
  • Department of Medical Physics, University of Wisconsin - Madison, Madison, Wisconsin, USA.
  • Engineering and Physical Sciences, University of Surrey, Guildford, UK.
  • Medical Radiation Physics, National Physical Laboratory, Teddington, UK.

ORCID-Profile: Andries N Schreuder

Zitieren

APA

Kissick, M. W., Panaino, C., Criscuolo, A., Hayes, J., Hoffman, C., Mackie, R., & Schreuder, A. N. (2024). Calculation method for novel upright CT scanner isodose curves. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14377

Vancouver

Kissick MW, Panaino C, Criscuolo A, Hayes J, Hoffman C, Mackie R, et al. Calculation method for novel upright CT scanner isodose curves. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14377

BibTeX
@article{Kissick2024,
  author = {Kissick, Michael W and Panaino, Costanza and Criscuolo, Anthony and Hayes, John and Hoffman, Carson and Mackie, Rockwell and Schreuder, Andries N},
  title = {Calculation method for novel upright CT scanner isodose curves},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14377}
}
Kostenlos starten →