A model that predicts a real-time tumour surface using intra-treatment skin surface and end-of-expiration and end-of-inhalation planning CT images
British Journal of Radiology · 2024
DOI: 10.1093/bjr/tqae067 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Radiology, 2024. Verfasst von Ziwen Wei und 9 weiteren Autor:innen. Beteiligte Einrichtung: Anhui Province Key Laboratory of Medical Physics and Technology, Institute of Health and Medical Technology, Hefei Institutes of Physical Science, He…. Hauptthemen laut MEDLINE: Lung Neoplasms, Four-Dimensional Computed Tomography, Skin. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Shandong Province, Anhui Province Funds for Distinguished Young Scientists.
Abstract (Originalsprache)
Objectives
To develop a mapping model between skin surface motion and internal tumour motion and deformation using end-of-exhalation (EOE) and end-of-inhalation (EOI) 3D CT images for tracking lung tumours during respiration.
Methods
Before treatment, skin and tumour surfaces were segmented and reconstructed from the EOE and the EOI 3D CT images. A non-rigid registration algorithm was used to register the EOE skin and tumour surfaces to the EOI, resulting in a displacement vector field that was then used to construct a mapping model. During treatment, the EOE skin surface was registered to the real-time, yielding a real-time skin surface displacement vector field. Using the mapping model generated, the input of a real-time skin surface can be used to calculate the real-time tumour surface. The proposed method was validated with and without simulated noise on 4D CT images from 15 patients at Léon Bérard Cancer Center and the 4D-lung dataset.
Results
The average centre position error, dice similarity coefficient (DSC), 95%-Hausdorff distance and mean distance to agreement of the tumour surfaces were 1.29 mm, 0.924, 2.76 mm, and 1.13 mm without simulated noise, respectively. With simulated noise, these values were 1.33 mm, 0.920, 2.79 mm, and 1.15 mm, respectively.
Conclusions
A patient-specific model was proposed and validated that was constructed using only EOE and EOI 3D CT images and real-time skin surface images to predict internal tumour motion and deformation during respiratory motion. Advances in knowledge The proposed method achieves comparable accuracy to state-of-the-art methods with fewer pre-treatment planning CT images, which holds potential for application in precise image-guided radiation therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Anhui Province Key Laboratory of Medical Physics and Technology, Institute of Health and Medical Technology, Hefei Institutes of Physical Science, Hefei Cancer Hospital, Chinese Academy of Sciences, Hefei 230031, P.R. C…
- Science Island Branch of the Graduate School, University of Science and Technology of China, Hefei 230026, Anhui, P.R. China.
- Department of Radiation Oncology, School of Medicine, Shandong University, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan 250117, Shandong, P.R.…
ORCID-Profile: Ziwen Wei, Junchao Qian
Förderung
- National Natural Science Foundation of China
- Natural Science Foundation of Shandong Province
- Anhui Province Funds for Distinguished Young Scientists
- Collaborative Innovation Program of Hefei Science Center
- China Postdoctoral Science Foundation
- Project of Postdoctoral Innovation of Shandong Province
Zitieren
APA
Wei, Z., Huang, X., Sun, A., Peng, L., Lou, Z., Hu, Z., Wang, H., Xing, L., Yu, J., & Qian, J. (2024). A model that predicts a real-time tumour surface using intra-treatment skin surface and end-of-expiration and end-of-inhalation planning CT images. British Journal of Radiology. https://doi.org/10.1093/bjr/tqae067
Vancouver
Wei Z, Huang X, Sun A, Peng L, Lou Z, Hu Z, et al. A model that predicts a real-time tumour surface using intra-treatment skin surface and end-of-expiration and end-of-inhalation planning CT images. British Journal of Radiology. 2024. doi:10.1093/bjr/tqae067
BibTeX
@article{Wei2024,
author = {Wei, Ziwen and Huang, Xiang and Sun, Aiming and Peng, Leilei and Lou, Zhixia and Hu, Zongtao and Wang, Hongzhi and Xing, Ligang and Yu, Jinming and Qian, Junchao},
title = {A model that predicts a real-time tumour surface using intra-treatment skin surface and end-of-expiration and end-of-inhalation planning CT images},
journal = {British Journal of Radiology},
year = {2024},
doi = {10.1093/bjr/tqae067}
}
EUVIMED