Investigating the impact of key algorithm parameters and patient‐specific factors on the accuracy of CT ventilation imaging

Jeremy Lim iD, John Kipritidis, Jeremy T Booth iD, Paul J Keall, Hilary L Byrne iD

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70610 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Jeremy Lim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Image X Institute, Faculty of Medicine and Health, The University of Sydney, Sydney, Australia. Hauptthemen laut MEDLINE: Algorithms, Lung Neoplasms, Four-Dimensional Computed Tomography, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cancer Institute NSW Translational Program Grant 2019.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Computed Tomography Ventilation Imaging (CTVI) is an investigational technique that has its basis in functional lung avoidance radiotherapy. It offers a cost‐effective and accessible alternative to nuclear medicine imaging by generating lung ventilation maps from 4DCT or paired inhale/exhale breath‐hold CT (BHCT) scans. Despite over a decade of clinical validation, there is still no consensus on how algorithm parameters and patient‐specific factors influence CTVI accuracy. Further research is needed to understand CTVI's sensitivity to these variables and to standardize its implementation for clinical use.

Purpose

This study evaluates how key algorithm parameters and patient‐specific factors affect the accuracy of CTVI. Materials and methods CT ventilation images were generated from BHCT scans and compared to Galligas PET ventilation scans. The VESPIR toolkit was used to compute ventilation based on deformable image registration (DIR) evaluation of volume change (CTVI Jac ) or change in Hounsfield Unit (HU) value (CTVI HU ). CTVI accuracy was characterized as the voxel‐wise Spearman correlation (r S ) with Galligas PET. Algorithm parameters common to many CTVI implementations were investigated with a baseline determined from existing literature: lung segmentation threshold (−600 HU to −150 HU), DIR regularization parameter (λ = 0.05 to 100), and smoothing filter diameter (0 voxels to 9 voxels). Robust parameter ranges were defined as those yielding r S within 10% of the maximum cohort average observed through parameter variation, and no negative Jacobian values for the registration. Patient‐specific lung volume and density metrics were also analyzed to explain inter‐patient variability in CTVI accuracy.

Results

The correlation between CTVI and Galligas PET was demonstrated to be robust within identified parameter ranges: lung segmentation threshold −600 HU to −150 HU for CTVI Jac and CTVI HU , DIR regularization parameter (λ) 1.25 to 5 for CTVI Jac and CTVI HU , and smoothing filter diameter 0 to 9 voxels for CTVI Jac and 7 to 9 voxels for CTVI HU . No significant correlation was found between the accuracy of CTVI Jac and any patient‐specific lung volume or density parameters. Significant correlations were found between the accuracy of CTVI HU and the percentage change in lung volume during inspiration ( r = 0.72, p < 0.01) and the lung volume in the exhale phase ( r = −0.63, p < 0.01). The correlation between CTVI Jac and CTVI HU was found to be strongly correlated to CTVI accuracy.

Conclusions

CTVI accuracy was relatively stable across the range of parameter values tested with no strong indication of the need for patient‐specific parameter sets. Patient‐specific differences appear to be a driving factor for inter‐patient variability in CTVI accuracy as parameter selection alone was insufficient to explain the variability. The strong association of CTVI Jac and CTVI HU agreement and CTVI accuracy suggests that CTVI Jac and CTVI HU agreement is a useful predictor of CTVI accuracy and quality metric for parameter optimization.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsLung NeoplasmsFour-Dimensional Computed TomographyRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedTomography, X-Ray ComputedHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

CT ventilation imagingalgorithm parameterssensitivity analysis

Autor:innen und Institutionen

  • Image X Institute, Faculty of Medicine and Health, The University of Sydney, Sydney, Australia.
  • Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, Australia.
  • Institute of Medical Physics, School of Physics, University of Sydney, Sydney, Australia.
  • Faculty of Medicine and Health, The University of Sydney, Sydney, Australia.

ORCID-Profile: Jeremy Lim, Jeremy T Booth, Hilary L Byrne

Förderung

  • Cancer Institute NSW Translational Program Grant 2019

Zitieren

APA

Lim, J., Kipritidis, J., Booth, J. T., Keall, P. J., & Byrne, H. L. (2026). Investigating the impact of key algorithm parameters and patient‐specific factors on the accuracy of CT ventilation imaging. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70610

Vancouver

Lim J, Kipritidis J, Booth JT, Keall PJ, Byrne HL. Investigating the impact of key algorithm parameters and patient‐specific factors on the accuracy of CT ventilation imaging. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70610

BibTeX
@article{Lim2026,
  author = {Lim, Jeremy and Kipritidis, John and Booth, Jeremy T and Keall, Paul J and Byrne, Hilary L},
  title = {Investigating the impact of key algorithm parameters and patient‐specific factors on the accuracy of CT ventilation imaging},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70610}
}
Kostenlos starten →