Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory

Toufik Medjadj iD, Mehenna Arib, Silia Hayoune iD

Biomedical Physics & Engineering Express · 2025

DOI: 10.1088/2057-1976/add34f ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, 2025. Verfasst von Toufik Medjadj und 2 weiteren Autor:innen. Beteiligte Einrichtung: Nuclear Research Center of Algiers, 02 Boulevard Frantz Fanon, PO Box 399, Alger RP, Alger, Algeria. Hauptthemen laut MEDLINE: Radiometry, Radiotherapy, Laboratories. Die Arbeit nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We aimed to design small fields collimators made of lead blocks, with diameters of 40 mm, 30 mm, 20 mm, and 10 mm, for the calibration of small volume radiotherapy ionization chambers using the Algerian Secondary Standard Dosimetry Laboratory (SSDL) ELDORADO 78 60 Co calibration unit. The radiation properties of the designed collimators were assessed using GAFChromic EBT3 dosimetry film (Ashland, USA) and two PTW small volume ionization chambers (PTW, Freiburg, Germany). Additionally, the photons spectra for the four different fields were also investigated using Monte Carlo (MC) method. The MC method was also utilized to define the dimensions of the collimators that achieve the desired field sizes at the reference distance of 80 cm. We simulated the photon beam of the 60 Co source from the irradiation unit to estimate the influencing quantities including field size and beam attenuation. EBT3 film, PTW Pinpoint and PTW Semiflex ionization chambers were used to measure the beam profiles, the output factors (OF), and the percentage depth doses (PDD) for the designed collimators. These measurements were performed at a depth of 5 cm and a source-to-surface distance of 80 cm. The beam profiles simulated with MC method and experimentally measured were found to be agreed within 3%/1 mm gamma criteria. The ionization chambers underestimate the OFs in comparison to EBT3 film for the four field sizes. The PDD curves were measured by the Pinpoint chamber and calculated using MC simulation. Significant differences were noted between the curves of the 20 and 10 mm collimators compared to those of the 40 mm and 30 mm ones. The curves were in agreement with the relative simulated data. Custom small-field collimators were successfully produced and characterized, which will facilitate studies on the variation and stability of calibration coefficients for small volume detectors operating within small-fields at our SSDL.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiometryRadiotherapyLaboratoriesCalibrationMonte Carlo MethodEquipment DesignRadiotherapy DosageHumans

Stichwörter der Autor:innen

Monte Carlo simulationSSDLcobalt-60radiotherapy calibrationsmall field collimators

Autor:innen und Institutionen

  • Nuclear Research Center of Algiers, 02 Boulevard Frantz Fanon, PO Box 399, Alger RP, Alger, Algeria.
  • King Faisal Specialist Hospital and Research Centre, PO Box 3354, Riyadh, 11211, Saudi Arabia.

ORCID-Profile: Toufik Medjadj, Silia Hayoune

Zitieren

APA

Medjadj, T., Arib, M., & Hayoune, S. (2025). Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/add34f

Vancouver

Medjadj T, Arib M, Hayoune S. Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory. Biomedical Physics & Engineering Express. 2025. doi:10.1088/2057-1976/add34f

BibTeX
@article{Medjadj2025,
  author = {Medjadj, Toufik and Arib, Mehenna and Hayoune, Silia},
  title = {Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory},
  journal = {Biomedical Physics & Engineering Express},
  year = {2025},
  doi = {10.1088/2057-1976/add34f}
}
Kostenlos starten →