Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192 Ir sources (HDR)

Douglas Batista da Silva iD, Ademir Xavier da Silva iD, Carlos Frederico Estrada Alves, Luis Alexandre Gonçalves Magalhaes iD

Biomedical Physics & Engineering Express · 2024

DOI: 10.1088/2057-1976/ad35cd ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, 2024. Verfasst von Douglas Batista da Silva und 3 weiteren Autor:innen. Beteiligte Einrichtung: Nuclear Engineering Program, Federal University of Rio de Janeiro (UFRJ), Rio de Janeiro, Brazil. Hauptthemen laut MEDLINE: Computer Simulation, Radiometry, Iridium Radioisotopes. Die Arbeit nennt 18 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this study was to use computer simulation to analyze the impact of the aluminum fixing support on the Reference Air Kerma (RAK), a physical quantity obtained in a calibration system that was experimentally developed in the Laboratory of Radiological Sciences of the University of the State of Rio de Janeiro (LCR-UERJ). Correction factors due to scattered radiation and the geometry of the 192 Ir sources were also sought to be determined. The computational simulation was validated by comparing some parameters of the experimental results with the computational results. These parameters were: verification of the inverse square law of distance, determination of (RAKR), analysis of the source spectrum with and without encapsulation, and the sensitivity curve of the Sourcecheck 4PI ionization chamber response, as a function of the distance from the source along the axial axis, using the microSelectron-v2 (mSv2) and GammaMedplus (GMp) sources. Kerma was determined by activity in the Reference air, with calculated values of 1.725 × 10 −3 U. Bq −1 and 1.710 × 10 −3 U. Bq −1 for the ionization chamber NE 2571 and TN 30001, respectively. The expanded uncertainty for these values was 0.932% and 0.919%, respectively, for a coverage factor (k = 2). The correction factor due to the influence of the aluminum fixing support for measurements at 1 cm and 10 cm from the source was 0.978 and 0.969, respectively. The geometric correction factor of the sources was k sg = 1.005 with an expanded uncertainty of 0.7% for a coverage factor (k = 2). This value has a difference of approximately 0.2% compared to the experimental values.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computer SimulationRadiometryIridium RadioisotopesCalibrationHumansAirAluminumMonte Carlo MethodRadiation DosageBrachytherapyRadiotherapy DosageScattering, Radiation

Stichwörter der Autor:innen

computer simulationcorrection factorsexpanded uncertaintykerma rate

Substanzen

Iridium RadioisotopesAluminum

Autor:innen und Institutionen

  • Nuclear Engineering Program, Federal University of Rio de Janeiro (UFRJ), Rio de Janeiro, Brazil.
  • Department of Radiological Sciences, Rio de Janeiro State University (UERJ), Rio de Janeiro, Brazil.

ORCID-Profile: Douglas Batista da Silva, Ademir Xavier da Silva, Luis Alexandre Gonçalves Magalhaes

Zitieren

APA

da Silva, D. B., da Silva, A. X., Alves, C. F. E., & Magalhaes, L. A. G. (2024). Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192 Ir sources (HDR). Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/ad35cd

Vancouver

da Silva DB, da Silva AX, Alves CFE, Magalhaes LAG. Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192 Ir sources (HDR). Biomedical Physics & Engineering Express. 2024. doi:10.1088/2057-1976/ad35cd

BibTeX
@article{daSilva2024,
  author = {da Silva, Douglas Batista and da Silva, Ademir Xavier and Alves, Carlos Frederico Estrada and Magalhaes, Luis Alexandre Gonçalves},
  title = {Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192 Ir sources (HDR)},
  journal = {Biomedical Physics & Engineering Express},
  year = {2024},
  doi = {10.1088/2057-1976/ad35cd}
}
Kostenlos starten →