Initial characterization of a novel dual-robot orthovoltage radiotherapy system
Biomedical Physics & Engineering Express · 31.03.2025 · 11 (2) · 025057
DOI: 10.1088/2057-1976/adbcb2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 17.03.2025. Verfasst von Olivia Masella und 4 weiteren Autor:innen. Beteiligte Einrichtung: University of Victoria, 3800 Finnerty Road, Victoria, BC, Canada. Hauptthemen laut MEDLINE: Robotics, Radiotherapy. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Discovery Accelerator Supplement, Discovery Grant, Canada Research Chairs.
Abstract (Originalsprache)
Purpose
Adequate access to radiotherapy is a critical global concern affecting low-resource settings such as low- and middle-income countries and rural regions. We propose to reduce this disparity by developing a novel low-cost radiotherapy device that treats using non-coplanar techniques and a 225 kVp x-ray tube.
Methods
This novel device has been preliminarily characterized spectrally, via spectrometer measurements, dosimetrically, via percent depth dose curves and 2D profiles, and geometrically, via a coplanar star-shot. Dosimetric and geometric evaluations were then combined by performing a proof of workflow of the KOALA system. Monte Carlo simulations were run in TOPAS to validate dosimetric measurements and the proof of workflow measurement.
Results
Spectral results showed excellent agreement between measured and modelled spectra. Dose errors of < 2% were achieved for PDD curves. Full width at half maximum values for the 2D profiles were, on average, 0.95 mm higher in simulation compared to film. A star-shot test demonstrated the high geometrical accuracy of the system with a 0.3 mm diameter wobble circle. Finally, a mean absolute percent error of 5 ± 5% (1 σ ) was measured for the proof of workflow test.
Conclusions
This initial characterization showcased the strengths and weaknesses of the KOALA system, with excellent isocenter precision and depth dose accuracy while lacking dosimetric accuracy in the 2D profiles. Further improvements on the source-to-collimator distance and treatment couch material can be made to improve the accuracy of a Monte Carlo model of the KOALA system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Victoria, 3800 Finnerty Road, Victoria, BC, Canada.
- McGill University, 845 Sherbrooke Street West, Montreal, QC, Canada.
- Sirius Medicine LLC, Half Moon Bay, CA, United States of America.
ORCID-Profile: Olivia Masella, Sandhya Rottoo, Magdalena Bazalova-Carter
Förderung
- Discovery Accelerator Supplement
- Discovery Grant
- Canada Research Chairs
Zitieren
APA
Masella, O., Atkinson, J., Rottoo, S., Weil, M., & Bazalova-Carter, M. (2025). Initial characterization of a novel dual-robot orthovoltage radiotherapy system. Biomedical Physics & Engineering Express, 11(2), 025057. https://doi.org/10.1088/2057-1976/adbcb2
Vancouver
Masella O, Atkinson J, Rottoo S, Weil M, Bazalova-Carter M. Initial characterization of a novel dual-robot orthovoltage radiotherapy system. Biomedical Physics & Engineering Express. 2025;11(2):025057. doi:10.1088/2057-1976/adbcb2
BibTeX
@article{Masella2025,
author = {Masella, Olivia and Atkinson, Jacob and Rottoo, Sandhya and Weil, Michael and Bazalova-Carter, Magdalena},
title = {Initial characterization of a novel dual-robot orthovoltage radiotherapy system},
journal = {Biomedical Physics & Engineering Express},
year = {2025},
volume = {11},
number = {2},
pages = {025057},
doi = {10.1088/2057-1976/adbcb2}
}
EUVIMED