Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh

Uttam Pyakurel, Arthur W Redgate iD, Carolyn A MacDonald iD, Jonathan C Petruccelli iD

Biomedical Physics & Engineering Express · 30.05.2025 · 11 (3) · 035024

DOI: 10.1088/2057-1976/adcd7d ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 30.04.2025. Verfasst von Uttam Pyakurel und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University at Albany, 1400, Washington Ave, Albany, NY, 12222, United States of America. Hauptthemen laut MEDLINE: Signal-To-Noise Ratio. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Phase differences imparted by tissue are significantly larger than attenuation differences. In addition, small angle scatter from tissue microstructure can provide a dark field signal that is complementary to attenuation and phase. Unfortunately, the low spatial coherence of clinical sources reduces phase and dark field contrast. Our method structures the beam with a single low-cost wire mesh that does not need precise alignment and relaxes the coherence requirement on the source. In addition, focusing polycapillary optics, which can be permanently attached to sources, are employed to allow for the use of high-power primary sources by increasing the phase signal after the focus. However, the coarseness of the mesh reduces the phase and dark field signal-to-noise ratio (SNR) compared with grating-based techniques, so optimization of the phase and dark-field SNR is an important consideration. Here, we consider the impact on the SNR of the distances between the mesh and the source and detector, and of x-ray tube voltages, to optimize the system.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Signal-To-Noise RatioAlgorithmsEquipment DesignPhantoms, ImagingX-Rays

Stichwörter der Autor:innen

computational imagingphase contrastx-ray dark fieldx-ray imaging

Autor:innen und Institutionen

  • Department of Physics, University at Albany, 1400, Washington Ave, Albany, NY, 12222, United States of America.

ORCID-Profile: Arthur W Redgate, Carolyn A MacDonald, Jonathan C Petruccelli

Zitieren

APA

Pyakurel, U., Redgate, A. W., MacDonald, C. A., & Petruccelli, J. C. (2025). Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh. Biomedical Physics & Engineering Express, 11(3), 035024. https://doi.org/10.1088/2057-1976/adcd7d

Vancouver

Pyakurel U, Redgate AW, MacDonald CA, Petruccelli JC. Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh. Biomedical Physics & Engineering Express. 2025;11(3):035024. doi:10.1088/2057-1976/adcd7d

BibTeX
@article{Pyakurel2025,
  author = {Pyakurel, Uttam and Redgate, Arthur W and MacDonald, Carolyn A and Petruccelli, Jonathan C},
  title = {Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh},
  journal = {Biomedical Physics & Engineering Express},
  year = {2025},
  volume = {11},
  number = {3},
  pages = {035024},
  doi = {10.1088/2057-1976/adcd7d}
}
Kostenlos starten →