EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dual-Readout Self-Resetting CMOS Image Sensor for Resolving Sub-Percent Optical Contrast in Biomedical Imaging

Kiyotaka Sasagawa iD, Subaru Iwaki, Kenji Morimoto, Ryoma Okada iD, Hironari Takehara iD, Makito Haruta iD, Hiroyuki Tashiro iD, Jun Ohta iD

Sensors · 2026

DOI: 10.3390/s26041396 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2026. Verfasst von Kiyotaka Sasagawa und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Materials Science, Graduate School of Science and Technology, Nara Institute of Science and Technology, 8916-5 Takayama, Ikoma 630-0192, …. Hauptthemen laut MEDLINE: Semiconductors. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Uehara Memorial Foundation, Terumo (Japan), Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We report a dual-readout self-resetting CMOS image sensor that achieves a signal-to-noise ratio (SNR) exceeding 70 dB and resolves sub-percent optical contrast variations by effectivly suppressing reset artifacts. The proposed sensor employs a Dual-Readout architecture with two independent scanners operating with a temporal offset; while one readout system is in the self-reset “dead time”, the other remains active, thereby physically ensuring continuous data acquisition. To minimize pixel area while achieving high reconstruction accuracy, a minimum frame-to-frame difference algorithm is utilized for signal restoration without requiring in-pixel counters. A prototype chip fabricated in a 0.35-μm process demonstrated SNR characteristics near the shot-noise limit, with a peak SNR exceeding 70 dB. Vascular phantom experiments using a carbon black suspension successfully visualized ±0.25% contrast fluctuations—dynamic signals previously undetectable by conventional sensors. This device provides a powerful platform for high-precision bio-imaging applications, including brain surface blood flow monitoring, where both wide dynamic range and high SNR are essential.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SemiconductorsSignal-To-Noise RatioAlgorithmsPhantoms, ImagingHumansImage Processing, Computer-AssistedEquipment Design

Stichwörter der Autor:innen

CMOS image sensorartifact reductionbiomedical imagingdual-readout architecturehigh dynamic rangemicrocirculationself-resetting pixelsignal-to-noise ratio

Autor:innen und Institutionen

  • Division of Materials Science, Graduate School of Science and Technology, Nara Institute of Science and Technology, 8916-5 Takayama, Ikoma 630-0192, Nara, Japan.
  • Medilux Research Center, Nara Institute of Science and Technology, 8916-5 Takayama, Ikoma 630-0192, Nara, Japan.
  • Institute for Research Initiatives, Nara Institute of Science and Technology, 8916-5 Takayama, Ikoma 630-0192, Nara, Japan.
  • Department of Opto-Electronic System Engineering, Chitose Institute of Science and Technology, 758-65 Bibi, Chitose 066-8655, Hokkaido, Japan.
  • Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku 812-8582, Fukuoka, Japan.

ORCID-Profile: Kiyotaka Sasagawa, Ryoma Okada, Hironari Takehara, Makito Haruta, Hiroyuki Tashiro, Jun Ohta

Förderung

  • Uehara Memorial Foundation
  • Terumo (Japan)
  • Japan Society for the Promotion of Science
  • Terumo Life Science Foundation
  • Japan Society for the Promotion of Science KAKENHI

Zitieren

APA

Sasagawa, K., Iwaki, S., Morimoto, K., Okada, R., Takehara, H., Haruta, M., Tashiro, H., & Ohta, J. (2026). Dual-Readout Self-Resetting CMOS Image Sensor for Resolving Sub-Percent Optical Contrast in Biomedical Imaging. Sensors. https://doi.org/10.3390/s26041396

Vancouver

Sasagawa K, Iwaki S, Morimoto K, Okada R, Takehara H, Haruta M, et al. Dual-Readout Self-Resetting CMOS Image Sensor for Resolving Sub-Percent Optical Contrast in Biomedical Imaging. Sensors. 2026. doi:10.3390/s26041396

BibTeX
@article{Sasagawa2026,
  author = {Sasagawa, Kiyotaka and Iwaki, Subaru and Morimoto, Kenji and Okada, Ryoma and Takehara, Hironari and Haruta, Makito and Tashiro, Hiroyuki and Ohta, Jun},
  title = {Dual-Readout Self-Resetting CMOS Image Sensor for Resolving Sub-Percent Optical Contrast in Biomedical Imaging},
  journal = {Sensors},
  year = {2026},
  doi = {10.3390/s26041396}
}