Evaluating the relationship between noise reduction and object thickness in scatter correction processing
Radiation Protection Dosimetry · 2025
DOI: 10.1093/rpd/ncaf095 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2025. Verfasst von Hiroki Tsuchiya und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Hospital of University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu-shi, Fukuoka 807-…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Polymethyl Methacrylate, Scattering, Radiation, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit nennt 21 Literaturangaben.
Abstract (Originalsprache)
Physical grids (PGs) are used in portable bedside radiography to reduce scattered X-rays that reach the detector. Because the use of PGs degrades image quality due to oblique incidence of X-rays or deviation of the center line on the grid, scatter correction (SC) processing is often used instead of PGs. This study aimed to evaluate the relationship between object thickness and noise reduction in SC processing. Polymethyl methacrylate (PMMA) phantoms of different thicknesses (16, 20, 26, and 30 cm) were used as objects. The Burger phantom was placed at the center of the thickness direction of the PMMA phantoms. Images were acquired with PGs, without PGs, and with SC processing which consists of contrast improvement processing and noise reduction processing. Contrast and noise were evaluated. Contrast of images without PG improved when SC processing was used. However, the effect of noise reduction process decreased with increase in the thickness of PMMA, and significant decrease was observed when the thickness of PMMA was ≥26 cm. In SC processing, the effect of noise reduction decreased with the increase in object thickness. These findings suggest that the applicability of SC processing may be limited, particularly in large patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Hospital of University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu-shi, Fukuoka 807-8555, Japan.
- Department of Radiological Science, Faculty of Health Sciences, Junshin Gakuen University, 1-1-1 Chikushigaoka, Minami-ku, Fukuoka 815-8510, Japan.
ORCID-Profile: Hiroki Tsuchiya
Zitieren
APA
Tsuchiya, H., Takaki, T., Ota, S., Fujita, Y., & Murakami, S. (2025). Evaluating the relationship between noise reduction and object thickness in scatter correction processing. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncaf095
Vancouver
Tsuchiya H, Takaki T, Ota S, Fujita Y, Murakami S. Evaluating the relationship between noise reduction and object thickness in scatter correction processing. Radiation Protection Dosimetry. 2025. doi:10.1093/rpd/ncaf095
BibTeX
@article{Tsuchiya2025,
author = {Tsuchiya, Hiroki and Takaki, Takeshi and Ota, Satoru and Fujita, Yuki and Murakami, Seiichi},
title = {Evaluating the relationship between noise reduction and object thickness in scatter correction processing},
journal = {Radiation Protection Dosimetry},
year = {2025},
doi = {10.1093/rpd/ncaf095}
}
EUVIMED