Dosimetry and Monte Carlo modelling of the Papillon+ contact X-ray brachytherapy device
Brachytherapy · 2024
DOI: 10.1016/j.brachy.2024.06.003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Brachytherapy, 2024. Verfasst von Dries Colson und 10 weiteren Autor:innen. Beteiligte Einrichtung: Hasselt University, Faculty of Engineering Technology, Nuclear Technological Centre (NuTeC), Hasselt, Belgium. Hauptthemen laut MEDLINE: Monte Carlo Method, Brachytherapy, Radiotherapy Dosage. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 33 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: This study aimed to develop and validate a Monte Carlo (MC) model for the Papillon+ contact x-ray brachytherapy (CXB) device, producing 50 kilovolt (kV) X-rays, specifically focusing on its application with a 25 mm diameter rectal applicator for contact therapy. MATERIAL AND METHODS: The validation process involved depth dose and transverse dose profile measurements using EBT3 gafchromic films positioned in a plastic water low energy range phantom. The half-value layer (HVL) was further measured and derived from the simulated X-ray spectra. RESULTS: Excellent agreement within ±2% was achieved between the measured and simulated on-axis depth dose curves for the 25 mm rectal applicator. Transverse dose profile measurements showed a high level of agreement between the simulation and measurements, on average 3.1% in contact with the applicator at the surface of the phantom and on average 1.7% at 10 mm depth. A close agreement within 5.5% was noticed concerning the HVL between the measurement and simulation. The simulated gamma spectra and 2D-dose distribution demonstrated a soft X-ray energy spectrum and a uniform dose distribution in contact with the applicator. CONCLUSIONS: An MC model was successfully developed for the Papillon+ eBT device with a 25 mm diameter rectal applicator. The validated model, with its demonstrated accuracy in depth dose and transverse dose profile simulations, is a valuable tool for quality assurance and patient safety and, in a later phase, may be used for treatment planning, dose calculations and tissue inhomogeneity corrections.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Hasselt University, Faculty of Engineering Technology, Nuclear Technological Centre (NuTeC), Hasselt, Belgium.
- Catharina Hospital, Department of Radiation Oncology, Eindhoven, The Netherlands.
- Department of Radiation Oncology (MAASTRO), GROW Research Institute for Oncology and Reproduction, Maastricht University Medical Center, the Netherlands.
- Institute for Research on Cancer and Aging of Nice (IRCAN), CNRS, INSERM, Université Côte d'Azur, Nice, France.
- Hasselt University, Faculty of Engineering Technology, Nuclear Technological Centre (NuTeC), Hasselt, Belgium. Electronic address: brigitte.reniers@uhasselt.be.
Zitieren
APA
Colson, D., Yalvac, B., Weterings, J., Verrijssen, A. S., van Haaren, P., Bellezzo, M., Berbée, M., Van Limbergen, E. J., Croce, O., Verhaegen, F., & Reniers, B. (2024). Dosimetry and Monte Carlo modelling of the Papillon+ contact X-ray brachytherapy device. Brachytherapy. https://doi.org/10.1016/j.brachy.2024.06.003
Vancouver
Colson D, Yalvac B, Weterings J, Verrijssen AS, van Haaren P, Bellezzo M, et al. Dosimetry and Monte Carlo modelling of the Papillon+ contact X-ray brachytherapy device. Brachytherapy. 2024. doi:10.1016/j.brachy.2024.06.003
BibTeX
@article{Colson2024,
author = {Colson, Dries and Yalvac, Burak and Weterings, Jan and Verrijssen, An-Sofie and van Haaren, Paul and Bellezzo, Murillo and Berbée, Maaike and Van Limbergen, Evert J and Croce, Olivier and Verhaegen, Frank and Reniers, Brigitte},
title = {Dosimetry and Monte Carlo modelling of the Papillon+ contact X-ray brachytherapy device},
journal = {Brachytherapy},
year = {2024},
doi = {10.1016/j.brachy.2024.06.003}
}
EUVIMED