Effects of count rate performance on image quality in an LAFOV PET/CT system
Physics in Medicine & Biology · 2026
DOI: 10.1088/1361-6560/aea301 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2026. Verfasst von Nanna Overbeck und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Physiology and Nuclear Medicine, Copenhagen University Hospital-Rigshospitalet, Copenhagen, Denmark. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The development of Long Axial Field-of-View positron emission tomography combined with computed tomography warrants reconsideration of the administered activity required for clinical imaging. This study investigated the relationship between body size, administered activity ( 18 F), and image quality (IQ) using two patient-like phantoms representing normal (78.5 kg) and obese (194.3 kg) body profiles. The phantoms were scanned in selected five-minute periods within a total duration of 27 and 21 h, respectively, following the decay of the activity. Analysis of the trues, scatter, and randoms demonstrated a close to linear propagation of the trues with activity up to a point where saturation was reached. The normal body profile phantom achieved higher signal-to-noise ratios (SNRs), reaching system saturation at a lower activity (680 MBq) than the large phantom, representing the obese body profile (970 MBq). The SNR 2 showed a nonlinear correlation with the noise equivalent count rate, whereas SNR 2 had a linear correlation with the trues. The study demonstrated the count rate capabilities up to and beyond the saturation limit. Furthermore, we illustrated the possibilities of scanning patients with ultralow dose administration in the range of 0.03–0.07 MBq kg −1 body weight. However, achieving robust clinical diagnostic quality—particularly for low-contrast lesions—at these activity levels requires further clinical investigation. Utilising these ultralow activity concentrations requires extending the scan duration to compensate for IQ, especially when examining obese patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Physiology and Nuclear Medicine, Copenhagen University Hospital-Rigshospitalet, Copenhagen, Denmark.
- Departement of Clinical Medicine, Copenhagen University, Copenhagen, Denmark.
ORCID-Profile: Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm, Flemming Littrup Andersen, Thomas Lund Andersen
Zitieren
APA
Overbeck, N., Czeluśniak, Z., Holm, S., Henriksen, J. V., Andersen, F. L., & Andersen, T. L. (2026). Effects of count rate performance on image quality in an LAFOV PET/CT system. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/aea301
Vancouver
Overbeck N, Czeluśniak Z, Holm S, Henriksen JV, Andersen FL, Andersen TL. Effects of count rate performance on image quality in an LAFOV PET/CT system. Physics in Medicine & Biology. 2026. doi:10.1088/1361-6560/aea301
BibTeX
@article{Overbeck2026,
author = {Overbeck, Nanna and Czeluśniak, Zuzanna and Holm, Søren and Henriksen, Julie V and Andersen, Flemming Littrup and Andersen, Thomas Lund},
title = {Effects of count rate performance on image quality in an LAFOV PET/CT system},
journal = {Physics in Medicine & Biology},
year = {2026},
doi = {10.1088/1361-6560/aea301}
}
EUVIMED