Investigation of optimal settings for deviceless respiratory synchronization in PET/CT examinations: effects of different reconstructions on image quality
Radiological Physics and Technology · 2025
DOI: 10.1007/s12194-025-00964-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiological Physics and Technology, 2025. Verfasst von Naoto Mori und 4 weiteren Autor:innen. Beteiligte Einrichtung: Section of Radiological Technology, Department of Medical Technology, Asahikawa Medical University Hospital, 2-1-1-1 Midorigaoka Higashi, Asahikawa, …. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted, Respiration. Die Arbeit nennt 20 Literaturangaben.
Abstract (Originalsprache)
Positron emission tomography (PET) images can be compromised by respiratory motion, leading to a decreased standardized uptake value (SUV) of the lesion and overestimation of the metabolic tumor volume (MTV). This study aimed to determine the optimal settings for auto-gating, a deviceless respiratory synchronization technique, using advanced intelligent clear-IQ engines (AiCE) and clear adaptive low-noise method (CaLM) reconstruction conditions. We performed phantom and clinical studies on 57 patients with pulmonary lesions. We acquired images at various %count settings (nongated, 30%, 50%, and 70%) using both AiCE and CaLM. In each setting, we measured the SUVmax, SUVpeak, and MTV of the lesions and calculated and compared the contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) in the liver. Additionally, we visually assessed lesion blurring and image noise to confirm whether the quantitative evaluation was consistent with the visual evaluation. Considering our findings, the optimal auto-gating setting at an acquisition time of 180 s is 50% for the lower lobe in AiCE and for both the lower and middle lobes in CaLM.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Section of Radiological Technology, Department of Medical Technology, Asahikawa Medical University Hospital, 2-1-1-1 Midorigaoka Higashi, Asahikawa, 078-8510, Japan. naomori@asahikawa-med.ac.jp.
- Section of Radiological Technology, Department of Medical Technology, Asahikawa Medical University Hospital, 2-1-1-1 Midorigaoka Higashi, Asahikawa, 078-8510, Japan.
- Department of Radiology, Asahikawa Medical University, Asahikawa, Japan.
Zitieren
APA
Mori, N., Iwata, K., Uno, T., Uchibe, T., & Okizaki, A. (2025). Investigation of optimal settings for deviceless respiratory synchronization in PET/CT examinations: effects of different reconstructions on image quality. Radiological Physics and Technology. https://doi.org/10.1007/s12194-025-00964-5
Vancouver
Mori N, Iwata K, Uno T, Uchibe T, Okizaki A. Investigation of optimal settings for deviceless respiratory synchronization in PET/CT examinations: effects of different reconstructions on image quality. Radiological Physics and Technology. 2025. doi:10.1007/s12194-025-00964-5
BibTeX
@article{Mori2025,
author = {Mori, Naoto and Iwata, Kunihiro and Uno, Takahiro and Uchibe, Taku and Okizaki, Atsutaka},
title = {Investigation of optimal settings for deviceless respiratory synchronization in PET/CT examinations: effects of different reconstructions on image quality},
journal = {Radiological Physics and Technology},
year = {2025},
doi = {10.1007/s12194-025-00964-5}
}
EUVIMED