Optimization of a Q.Clear PET image reconstruction based on bayesian optimization in oncology and neurology applications
Physical and Engineering Sciences in Medicine · 2026
DOI: 10.1007/s13246-026-01776-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2026. Verfasst von Qariemah Azahar und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Imaging, Pusat Kanser Tun Abdullah Ahmad Badawi, Universiti Sains Malaysia, 13200, Kepala Batas, Penang, Malaysia. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Neurology, Positron Emission Tomography Computed Tomography, Positron-Emission Tomography. Die Arbeit nennt 33 Literaturangaben. Gefördert von: Fundamental Research Grant Scheme (FRGS).
Abstract (Originalsprache)
Bayesian Penalized Likelihood (BPL) reconstruction can improve image quality and quantitative accuracy in PET/CT imaging. This study aimed to determine the optimal Beta (β)-factor in Q.Clear reconstruction using Bayesian Optimization (BO) through phantoms and clinical evaluation. NEMA IQ and Hoffman Brain phantoms, along with clinical PET/CT images, were analyzed. Phantom images were reconstructed using β-factors (50-1000) across different acquisition times. Recovery coefficient (RC%) and contrast-to-noise ratio (CNR) were calculated. Clinical evaluation included one neurology and one oncology case as verification data, reconstructed using OSEM and Q.Clear at β-factors (100-1000), with full and reduced reconstruction times (neurology: 8 and 2 min; oncology: 15 and 7 min), respectively. Visual assessment was independently performed by two nuclear medicine physicians, and inter-rater agreement was analyzed using Krippendorff's alpha. In the NEMA IQ phantom, increasing β-factor resulted in decreased RC% but increased CNR. In the Hoffman Brain phantom, cortical regions showed stable RC% from mid to high β-factors, whereas subcortical regions were more sensitive, showing reduced RC%. CNR improved with higher β-factors and longer acquisition times. BO analysis indicates that smaller spheres achieved optimal performance at lower β-factors and longer acquisition times, while larger spheres tolerated higher β-factors (600-800). Optimal β-factors for cortical and subcortical brain regions were (400-600) and (100-300), respectively. Clinically, the highest scores were observed for β-factors (300-1000) in oncology, and (500-900) in neurology. Higher β-factors improved RC% and CNR stability in phantom studies. BO identified (500-900) as the optimal β-factors, consistent with clinical findings demonstrating high visual scores and strong inter-rater agreement, supporting its reliability of BO-optimized β-factors for PET/CT reconstruction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Imaging, Pusat Kanser Tun Abdullah Ahmad Badawi, Universiti Sains Malaysia, 13200, Kepala Batas, Penang, Malaysia.
- Unit Perubatan Nuklear, Pusat Kanser Tun Abdullah Ahmad Badawi, Universiti Sains Malaysia, 13200, Kepala Batas, Penang, Malaysia.
- School of Health Sciences, Universiti Sains Malaysia, 16150, Kubang Kerian, Kelantan, Malaysia.
- Department of Biomedical Imaging, Pusat Kanser Tun Abdullah Ahmad Badawi, Universiti Sains Malaysia, 13200, Kepala Batas, Penang, Malaysia. syahir.mansor@usm.my.
- Unit Perubatan Nuklear, Pusat Kanser Tun Abdullah Ahmad Badawi, Universiti Sains Malaysia, 13200, Kepala Batas, Penang, Malaysia. syahir.mansor@usm.my.
Förderung
- Fundamental Research Grant Scheme (FRGS)
Zitieren
APA
Azahar, Q., Hashim, H., Abdul Hamid, K., Xin, L., & Mansor, S. (2026). Optimization of a Q.Clear PET image reconstruction based on bayesian optimization in oncology and neurology applications. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-026-01776-6
Vancouver
Azahar Q, Hashim H, Abdul Hamid K, Xin L, Mansor S. Optimization of a Q.Clear PET image reconstruction based on bayesian optimization in oncology and neurology applications. Physical and Engineering Sciences in Medicine. 2026. doi:10.1007/s13246-026-01776-6
BibTeX
@article{Azahar2026,
author = {Azahar, Qariemah and Hashim, Hazlin and Abdul Hamid, Khadijah and Xin, Lyu and Mansor, Syahir},
title = {Optimization of a Q.Clear PET image reconstruction based on bayesian optimization in oncology and neurology applications},
journal = {Physical and Engineering Sciences in Medicine},
year = {2026},
doi = {10.1007/s13246-026-01776-6}
}