Assessing effects of low‐dose CT attenuation correction on quantification and image quality on a LAFOV PET system with a modified NEMA‐IQ phantom

Vanessa Nadig, Jane Mackewn, Georgios Krokos, Lewis C Davies, Daniel Darian, Sarah-May Gould, Gary J R Cook, Alexander Hammers, Andrew J Reader, Paul K Marsden

Medical Physics · 2026-11 · 53 (11)

DOI: 10.1002/mp.70706 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 01.11.2026. Verfasst von Vanessa Nadig und 9 weiteren Autor:innen. Beteiligte Einrichtung: Siemens Healthcare Limited, Camberley, UK. Hauptthemen laut MEDLINE: Radiation Dosage, Phantoms, Imaging, Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical Research Council (URKI Infrastructure Fund) / National PET Imaging Platform, National Institute for Health and Care Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Over the past decades, efforts have been directed toward increasing the sensitivity of PET systems to enable low‐dose PET imaging and shorter scan times. The Biograph Vision Quadra PET/CT scanner (Siemens Healthineers, Knoxville, TN, USA) combines an increased solid angle coverage with time of flight, enabling a significant reduction in administered activity. To fully benefit from low‐dose PET imaging, the dose of the CT scan should ideally also be reduced.

Purpose

The CT radiation dose can be reduced by lowering the tube voltage and exposure as well as by the use of tin filtration. Reducing the CT dose must be carefully aligned with any impact that changing CT image quality and quantitative metrics may have on the PET quantification, for example via linear attenuation coefficients. We evaluate the impact of low‐dose CT protocols on the PET quantification in soft‐tissue‐like and bone‐equivalent materials using a customized phantom.

Methods

The phantom was assembled out of a NEMA‐IQ phantom, several layers of saline and two bovine long bones, to simulate a patient positioned with arms down. A 10‐min PET scan with 47 different accompanying CT scans was acquired. The CT tube voltage and exposure were varied between , with and without tin filtration, and , respectively. The effect of the different CT parameters on the PET image quality and quantitative accuracy was assessed via several quantitative metrics, including standardized uptake value (SUV) and signal‐to‐noise ratio, along with the CT dose length product (DLP).

Results

The recorded DLP values suggest that using all tube voltages except combined with exposures in the range of , can result in dose savings of compared to a typical clinical protocol (, ). Compared to this typical protocol, HU differed more in bone ( difference) than in soft tissue ( difference), and more streak artefacts and photon starvation were seen when configuring low tube voltages and exposures, with an accompanying increase in CT noise. The observed artefacts were transferred with low intensity to the PET images via the attenuation map. The remained within of the of the typical clinical protocol for most tested protocols. The presence of artefacts in the CT images had a bigger impact on PET quantification in both soft tissue and bone ( difference of ) than any difference in HU. Protocols that resulted in artefacts showed an increased inaccuracy in attenuation correction and scatter estimation, as assessed via the “lung” residual error (up to increase).

Conclusions

Even for low‐dose CT protocols, PET quantification remained consistent in all NEMA‐IQ spheres for many materials. Beam hardening and photon starvation in the CT image are likely to occur in specific body regions such as hips, shoulders or the abdomen for an arms‐down scan and these effects had a bigger impact on the quantification than any difference in HU in both soft tissue and bone. This therefore should be considered when selecting suitable CT imaging parameters.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosagePhantoms, ImagingPositron Emission Tomography Computed TomographyImage Processing, Computer-AssistedTomography, X-Ray ComputedAnimalsSignal-To-Noise RatioCattle

Stichwörter der Autor:innen

LAFOVPET/CTlow‐dose imagingquantificationtin filtertotal‐body PET

Autor:innen und Institutionen

  • Siemens Healthcare Limited, Camberley, UK.
  • King's College London and Guy's & St Thomas' PET Centre, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Department of Biomedical Computing, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Department of Early Life Imaging, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.

Förderung

  • Medical Research Council (URKI Infrastructure Fund) / National PET Imaging Platform
  • National Institute for Health and Care Research

Zitieren

APA

Nadig, V., Mackewn, J., Krokos, G., Davies, L. C., Darian, D., Gould, S. M., Cook, G. J. R., Hammers, A., Reader, A. J., & Marsden, P. K. (2026). Assessing effects of low‐dose CT attenuation correction on quantification and image quality on a LAFOV PET system with a modified NEMA‐IQ phantom. Medical Physics, 53(11). https://doi.org/10.1002/mp.70706

Vancouver

Nadig V, Mackewn J, Krokos G, Davies LC, Darian D, Gould SM, et al. Assessing effects of low‐dose CT attenuation correction on quantification and image quality on a LAFOV PET system with a modified NEMA‐IQ phantom. Medical Physics. 2026;53(11). doi:10.1002/mp.70706

BibTeX
@article{Nadig2026,
  author = {Nadig, Vanessa and Mackewn, Jane and Krokos, Georgios and Davies, Lewis C and Darian, Daniel and Gould, Sarah-May and Cook, Gary J R and Hammers, Alexander and Reader, Andrew J and Marsden, Paul K},
  title = {Assessing effects of low‐dose CT attenuation correction on quantification and image quality on a LAFOV PET system with a modified NEMA‐IQ phantom},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {11},
  doi = {10.1002/mp.70706}
}
Kostenlos starten →