In Vitro Evaluation of a Device to Assess Acetabular Cup Position by Sagittal Plane Radiography
Veterinary and Comparative Orthopaedics and Traumatology · 01.03.2026 · 39 (02) · 075-080
DOI: 10.1055/a-2651-6297 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Veterinary and Comparative Orthopaedics and Traumatology, erschienen am 01.03.2026. Verfasst von W Michael Karlin und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Sciences, Cummings School of Veterinary Medicine at Tufts University, North Grafton, Massachusetts, United States. Hauptthemen laut MEDLINE: Acetabulum, Radiography, Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit nennt 12 Literaturangaben.
Abstract (Originalsprache)
The purpose of this study was to compare the calculated angle of lateral opening (ALO) and open-face version obtained using an acetabular cup position assessment device (CPAD) to the known ALO and open-face version of an acetabular component. The null hypothesis was that there would be no significant differences between the ALO and open-face versions using the CPAD compared with the known values. Radiographic images simulating a sagittal plane pelvis were obtained with a BFX™ acetabular component containing a CPAD mounted on a plexiglass jig with a dual-plane inclinometer. Images were obtained at an open-face version of 0 degrees with ALO at 30, 40, 45, 50, and 60 degrees, repeated in triplicate, and were randomized. Measurements of the CPAD bars were used to calculate the open-face version, and ALO measurements were compared using a paired t-test with a p-value set at p < 0.05. Agreement was assessed with a Lin's Concordance Correlation Coefficient. The calculated ALO was not significantly different from the known ALO (p = 0.608) and showed near-perfect correlation of the ALO with a Lin's Concordance Correlation Coefficient (R c value) of 0.994. The calculated open-face version was significantly different from the known version. The results demonstrate that the CPAD is an accurate method to determine ALO in a simulated sagittal plane pelvis radiographic view.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Clinical Sciences, Cummings School of Veterinary Medicine at Tufts University, North Grafton, Massachusetts, United States.
- Arizona Canine Orthopedics and Sports Medicine, Scottsdale, Arizona, United States.
Zitieren
APA
Karlin, W. M., Lirtzman, R., & Kowaleski, M. P. (2026). In Vitro Evaluation of a Device to Assess Acetabular Cup Position by Sagittal Plane Radiography. Veterinary and Comparative Orthopaedics and Traumatology, 39(02), 075-080. https://doi.org/10.1055/a-2651-6297
Vancouver
Karlin WM, Lirtzman R, Kowaleski MP. In Vitro Evaluation of a Device to Assess Acetabular Cup Position by Sagittal Plane Radiography. Veterinary and Comparative Orthopaedics and Traumatology. 2026;39(02):075-080. doi:10.1055/a-2651-6297
BibTeX
@article{Karlin2026,
author = {Karlin, W Michael and Lirtzman, Ross and Kowaleski, Michael P},
title = {In Vitro Evaluation of a Device to Assess Acetabular Cup Position by Sagittal Plane Radiography},
journal = {Veterinary and Comparative Orthopaedics and Traumatology},
year = {2026},
volume = {39},
number = {02},
pages = {075--080},
doi = {10.1055/a-2651-6297}
}
EUVIMED