Fachartikel in der Zeitschrift Radiography, 2023-07. Verfasst von A-M Sandbukt Johnsen und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Nuclear Medicine, Oslo University Hospital, Sognsvannsveien 20, 0372 Oslo, Norway; Faculty of Health Sciences, Department…. Hauptthemen laut MEDLINE: Algorithms, Tomography, X-Ray Computed. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
INTRODUCTION: This study investigated the impact of high matrix image reconstruction in combination with different reconstruction kernels and levels of iterative reconstructions on image quality in chest CT. METHODS: An anthropomorphic chest phantom (Kyoto Kagaku Co., Ltd., Kyoto, Japan), and a Catphan® 600 (The Phantom Laboratory, Greenwich, NY, USA) phantom were scanned using a dual source scanner. Standard institutional protocol with 512 × 512 matrix was used as a reference. Reconstructions were performed for 768 × 768 and 1024 × 1024 matrices and all possible combinations of three different kernels and five levels of iterative reconstructions were included. Signal difference to noise ratio (SdNR) and line pairs per cm (lp/cm) were manually measured. A Linear regression model was applied for objective image analysis (SdNR) and inter-and intra-reader agreement was given as Cohen's kappa for the visual image assessment. RESULTS: Matrix size did not have a significant impact on SdNR (p = 0.595). Kernel (p = 0.014) and ADMIRE level (p = 0.001) had a statistically significant impact on SdNR. The spatial resolution ranged from 7 lp/cm to 9 lp/cm. The highest spatial resolution was achieved using kernel Br64 and ADMIRE 1, 2 and 3 in both 768- and 1024-matrices, and with Br59 with ADMIRE 2 and 4 and 768-matrix, all visualizing 9 lp/cm. Both readers scored kernel Br59 highest, and the scoring increased with increasing levels of Iterative Reconstruction. CONCLUSION: Matrix size did not influence image quality, however, the choice of kernel and degree of IR had an impact on objective and visual image quality in 768 - and 1024-matrices, suggesting that increased degree of IR may improve diagnostic image quality in chest CT. IMPLICATIONS FOR PRACTICE: Image quality in CT of the lung may be improved by increasing the level of IR.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Nuclear Medicine, Oslo University Hospital, Sognsvannsveien 20, 0372 Oslo, Norway; Faculty of Health Sciences, Department of Life Sciences and Health, Oslo Metropolitan University, Pilestrede…
- Department of Radiology and Nuclear Medicine, Oslo University Hospital, Sognsvannsveien 20, 0372 Oslo, Norway. Electronic address: jkristia@ous-hf.no.
- Faculty of Health Sciences, Department of Life Sciences and Health, Oslo Metropolitan University, Pilestredet 48, 0130 Oslo, Norway. Electronic address: ousjobb@gmail.com.
Zitieren
APA
Sandbukt Johnsen, A. M., Fenn, J. M., Henning, M. K., & Hauge, I. H. (2023). Optimization of chest CT protocols based on pixel image matrix, kernels and iterative reconstruction levels – A phantom study. Radiography, 29(4), 752-759. https://doi.org/10.1016/j.radi.2023.05.005
Vancouver
Sandbukt Johnsen AM, Fenn JM, Henning MK, Hauge IH. Optimization of chest CT protocols based on pixel image matrix, kernels and iterative reconstruction levels – A phantom study. Radiography. 2023;29(4):752-759. doi:10.1016/j.radi.2023.05.005
BibTeX
@article{SandbuktJohnsen2023,
author = {Sandbukt Johnsen, A-M and Fenn, J M and Henning, M K and Hauge, I H},
title = {Optimization of chest CT protocols based on pixel image matrix, kernels and iterative reconstruction levels – A phantom study},
journal = {Radiography},
year = {2023},
volume = {29},
number = {4},
pages = {752--759},
doi = {10.1016/j.radi.2023.05.005}
}
EUVIMED