Impact of CT scan parameters on deformable image registration accuracy using deformable thorax phantom

Ryutaro Ikeda, Noriyuki Kadoya, Yujiro Nakajima, Shin Ishii, Takayuki Shibu, Keiichi Jingu

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.13917 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Ryutaro Ikeda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Tomography, X-Ray Computed. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The purpose of this study was to evaluate the deformable image registration (DIR) accuracy using various CT scan parameters with deformable thorax phantom. Our developed deformable thorax phantom (Dephan, Chiyoda Technol Corp, Tokyo, Japan) was used. The phantom consists of a base phantom, an inner phantom, and a motor‐derived piston. The base phantom is an acrylic cylinder phantom with a diameter of 180 mm, which simulates the chest wall. The inner phantom consists of deformable, 20 mm thick disk‐shaped sponges with 48 Lucite beads and 48 nylon cross‐wires which simulate the vascular and bronchial bifurcations of the lung. Peak‐exhale and peak‐inhale images of the deformable phantom were acquired using a CT scanner (Aquilion LB, TOSHIBA). To evaluate the impact of CT scan parameters on DIR accuracy, we used the four tube voltages (80, 100, 120, and 135 kV) and six reconstruction algorithms (FC11, FC13, FC15, FC41, FC44, and FC52). Intensity‐based DIR was performed between the two images using MIM Maestro (MIM software, Cleveland, USA). Fiducial markers (beads and cross‐wires) based target registration error (TRE) was used for quantitative evaluation of DIR. In case with different tube voltages, the range of average TRE were 4.44–5.69 mm (reconstruction algorithm: FC13). In case with different reconstruction algorithms, the range of average TRE were 4.26–4.59 mm (tube voltage: 120 kV). The TRE were differed by up to 3.0 mm (3.96–6.96 mm) depending on the combination of tube voltage and reconstruction algorithm. Our result indicated that CT scan parameters had moderate impact of TRE, especially for reconstruction algorithms for the deformable thorax phantom.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Processing, Computer-AssistedTomography, X-Ray ComputedHumansSoftwareAlgorithmsThoraxPhantoms, Imaging

Stichwörter der Autor:innen

deformable image registrationlungphysical phantomradiotherapytarget registration error

Autor:innen und Institutionen

  • Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan.
  • Department of Radiology, Japanese Red Cross Ishinomaki Hospital, Ishinomaki, Japan.
  • Department of Radiological Science, Komazawa University, Tokyo, Japan.

Zitieren

APA

Ikeda, R., Kadoya, N., Nakajima, Y., Ishii, S., Shibu, T., & Jingu, K. (2023). Impact of CT scan parameters on deformable image registration accuracy using deformable thorax phantom. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13917

Vancouver

Ikeda R, Kadoya N, Nakajima Y, Ishii S, Shibu T, Jingu K. Impact of CT scan parameters on deformable image registration accuracy using deformable thorax phantom. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13917

BibTeX
@article{Ikeda2023,
  author = {Ikeda, Ryutaro and Kadoya, Noriyuki and Nakajima, Yujiro and Ishii, Shin and Shibu, Takayuki and Jingu, Keiichi},
  title = {Impact of CT scan parameters on deformable image registration accuracy using deformable thorax phantom},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.13917}
}
Kostenlos starten →