EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Accuracy and safety of implant placement with a novel semi-autonomous robotic-assisted surgical system: A translational research study

Fan Yang, Jianping Chen, Linhong Wang, Yude Ding

The Journal of Prosthetic Dentistry · 2025-05 · 133 (5) · 1304-1314

DOI: 10.1016/j.prosdent.2024.10.026 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Prosthetic Dentistry, 2025-05. Verfasst von Fan Yang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Professor, Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospita…. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Dental Implantation, Endosseous. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Zhejiang Administration Bureau of Traditional Chinese Medicine, Zhejiang Chinese Medical University, Medical Science and Technology Project of Zhejiang Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

STATEMENT OF PROBLEM: Translational studies evaluating the feasibility, accuracy, and safety of semi-autonomous implant robots, from model test to animal experiment and clinical trial, are currently lacking. PURPOSE: The purpose of this study was to evaluate the accuracy and safety of a novel semi-autonomous robotic-assisted surgical system (sa-RASS) in implant placement. MATERIAL AND METHODS: A translational study was conducted to assess the application of the sa-RASS in a model test, an animal experiment, and a clinical controlled trial. The study included 45 resin models in the model test, 7 male beagle dogs in the animal experiment, and 60 participants who were recruited and randomly assigned to a freehand or a sa-RASS group in the clinical trial. The accuracy, surgical morbidity, complications, operator ratings of instrument safety, and satisfaction were recorded. Cone beam computed tomography data were used to evaluate deviations between planned and placed implants. The data on deviations were analyzed using the Mann-Whitney U test. A linear regression model was established to analyze the variations in the deviations (α=.05). RESULTS: The mean ±standard deviation of the platform, apex, and angulation deviations were 0.98 ±0.53 mm, 1.10 ±0.52 mm, and 1.45 ±0.60 degrees in the model test and 0.58 ±0.19, 0.59 ±0.22 mm, and 1.88 ±0.71 degrees in the animal experiment. In the clinical controlled study, the platform, apex, and angular deviations of the sa-RASS group and freehand group were 0.93 ±0.50 mm versus 1.45 ±0.86 mm (P .05). CONCLUSIONS: The sa-RASS was found to be more accurate than freehand placement, with high operational safety and low surgical morbidity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresDental Implantation, EndosseousDogsAnimalsMaleHumansTranslational Research, BiomedicalCone-Beam Computed TomographyFemaleAdult

Autor:innen und Institutionen

  • Professor, Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, PR China.
  • Attending, Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, PR China.
  • Associate Professor, Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, PR China.
  • Attending, Center for Plastic & Reconstructive Surgery, Department of Stomatology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, PR China. Electronic…

Förderung

  • Zhejiang Administration Bureau of Traditional Chinese Medicine
  • Zhejiang Chinese Medical University
  • Medical Science and Technology Project of Zhejiang Province

Zitieren

APA

Yang, F., Chen, J., Wang, L., & Ding, Y. (2025). Accuracy and safety of implant placement with a novel semi-autonomous robotic-assisted surgical system: A translational research study. The Journal of Prosthetic Dentistry, 133(5), 1304-1314. https://doi.org/10.1016/j.prosdent.2024.10.026

Vancouver

Yang F, Chen J, Wang L, Ding Y. Accuracy and safety of implant placement with a novel semi-autonomous robotic-assisted surgical system: A translational research study. The Journal of Prosthetic Dentistry. 2025;133(5):1304-1314. doi:10.1016/j.prosdent.2024.10.026

BibTeX
@article{Yang2025,
  author = {Yang, Fan and Chen, Jianping and Wang, Linhong and Ding, Yude},
  title = {Accuracy and safety of implant placement with a novel semi-autonomous robotic-assisted surgical system: A translational research study},
  journal = {The Journal of Prosthetic Dentistry},
  year = {2025},
  volume = {133},
  number = {5},
  pages = {1304--1314},
  doi = {10.1016/j.prosdent.2024.10.026}
}
Kostenlos starten →