Geometrically-derived density compensation function for gridding reconstruction of 2D and 3D non-Cartesian MRI data
Magnetic Resonance Imaging · 2026-07 · 130 · 110674
DOI: 10.1016/j.mri.2026.110674 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2026-07. Verfasst von Oluyemi Aboyewa und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Northwestern University, Evanston, IL, United States of America; Department of Radiology, Feinberg School of Me…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Imaging, Three-Dimensional, Brain, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: American Heart Association, Radiological Society of North America.
Abstract (Originalsprache)
PURPOSE Density compensation function (DCF) is essential for minimizing reconstruction errors in non-Cartesian MRI. This study aimed to extend a geometrically-derived DCF (gDCF), previously validated for 2D radial MRI, to arbitrary 2D and 3D non-Cartesian k-space sampling trajectories. METHODS In the gDCF framework, sample weights were analytically computed as the inverse sum of the overlap degrees between neighboring radial k-space samples, using a Nyquist sampling distance, ∆k. To generalize gDCF beyond radial trajectories, a direction-independent ∆k was introduced. Three approaches for redefining the sampling distance were evaluated: Nyquist distance (∆kNyq=1/Field-of-View), generalized FOV (∆kgen), and Beruling-Landau density condition (∆kBL). To account for modulation introduced by the convolution kernel used in the gridding reconstruction, a kernel correction was integrated into the gDCF formulation. The proposed gDCF methods were evaluated using numerical simulations on fully sampled 2D phantom and T1-weighted 3D brain MRI datasets. Performance was benchmarked against existing DCF methods, including ramp/rho-filter, Voronoi, Pipe, and LSQR, using radial, spiral, rosette, and cone k-space trajectories. Image quality was quantitatively assessed using root-mean-square error (RMSE) and structural-similarity index (SSIM) against the fully sampled references. RESULTS In 2D simulations, kernel-corrected gDCF achieved reconstruction accuracy (RMSE≤0.0190 and SSIM≥0.9756) comparable to Pipe (RMSE≤0.0189; SSIM≥0.9758) across all trajectories. In 3D simulations, kernel-corrected gDCFNyq outperformed the Rho-filter but remained less accurate than Pipe. Nevertheless, gDCFNyqwas ∼2× faster than Pipe at larger k-space matrix sizes. CONCLUSIONS This study demonstrated that gDCF could be extended to arbitrary non-Cartesian trajectories, providing accurate and computationally efficient density compensations for gridding-based MRI reconstruction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Northwestern University, Evanston, IL, United States of America; Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, IL, United States of America.
- Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, IL, United States of America. Electronic address: kyungpyo.hong@northwestern.edu.
Förderung
- American Heart Association
- Radiological Society of North America
Zitieren
APA
Aboyewa, O., Kim, D., & Hong, K. (2026). Geometrically-derived density compensation function for gridding reconstruction of 2D and 3D non-Cartesian MRI data. Magnetic Resonance Imaging, 130, 110674. https://doi.org/10.1016/j.mri.2026.110674
Vancouver
Aboyewa O, Kim D, Hong K. Geometrically-derived density compensation function for gridding reconstruction of 2D and 3D non-Cartesian MRI data. Magnetic Resonance Imaging. 2026;130:110674. doi:10.1016/j.mri.2026.110674
BibTeX
@article{Aboyewa2026,
author = {Aboyewa, Oluyemi and Kim, Daniel and Hong, KyungPyo},
title = {Geometrically-derived density compensation function for gridding reconstruction of 2D and 3D non-Cartesian MRI data},
journal = {Magnetic Resonance Imaging},
year = {2026},
volume = {130},
pages = {110674},
doi = {10.1016/j.mri.2026.110674}
}
EUVIMED