Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 19.03.2024. Verfasst von Matthew J Middione und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Stanford University, Stanford, California. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Brain, Algorithms. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIMH NIH HHS, NIBIB NIH HHS, National Institute of Mental Health.
Abstract (Originalsprache)
Purpose
To evaluate the use of pre‐excitation gradients for eddy current–nulled convex optimized diffusion encoding (Pre‐ENCODE) to mitigate eddy current–induced image distortions in diffusion‐weighted MRI (DWI).
Methods
DWI sequences using monopolar (MONO), ENCODE, and Pre‐ENCODE were evaluated in terms of the minimum achievable echo time (TE) and eddy current‐induced image distortions using simulations, phantom experiments, and in vivo DWI in volunteers ().
Results
Pre‐ENCODE provided a shorter TE than MONO (71.0 17.7ms vs. 77.6 22.9ms) and ENCODE (71.0 17.7ms vs. 86.2 14.2ms) in 100 of the simulated cases for a commercial 3T MRI system with b ‐values ranging from 500 to 3000 s/mm and in‐plane spatial resolutions ranging from 1.0 to 3.0mm. Image distortion was estimated by intravoxel signal variance between diffusion encoding directions near the phantom edges and was significantly lower with Pre‐ENCODE than with MONO (10.1 vs. 22.7, ) and comparable to ENCODE (10.1 vs. 10.4, ). In vivo measurements of apparent diffusion coefficients were similar in global brain pixels (0.37 [0.28,1.45] mm/s vs. 0.38 [0.28,1.45]mm/s, ) and increased in edge brain pixels (0.80 [0.17,1.49] mm/s vs. 0.70 [0.18,1.48]mm/s, ) for MONO compared to Pre‐ENCODE.
Conclusion
Pre‐ENCODE mitigated eddy current–induced image distortions for diffusion imaging with a shorter TE than MONO and ENCODE.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Stanford University, Stanford, California.
- Department of Electrical Engineering, Stanford University, Stanford, California.
ORCID-Profile: Matthew J Middione, Michael Loecher, Xiaozhi Cao, Kawin Setsompop, Daniel B Ennis
Förderung
- NIMH NIH HHS
- NIBIB NIH HHS
- National Institute of Mental Health
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Middione, M. J., Loecher, M., Cao, X., Setsompop, K., & Ennis, D. B. (2024). Pre‐excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre‐ENCODE). Magnetic Resonance in Medicine, 92(2), 573-585. https://doi.org/10.1002/mrm.30068
Vancouver
Middione MJ, Loecher M, Cao X, Setsompop K, Ennis DB. Pre‐excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre‐ENCODE). Magnetic Resonance in Medicine. 2024;92(2):573-585. doi:10.1002/mrm.30068
BibTeX
@article{Middione2024,
author = {Middione, Matthew J and Loecher, Michael and Cao, Xiaozhi and Setsompop, Kawin and Ennis, Daniel B},
title = {Pre‐excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre‐ENCODE)},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {92},
number = {2},
pages = {573--585},
doi = {10.1002/mrm.30068}
}
EUVIMED