Accuracy assessment (trueness and precision) of a confocal based intraoral scanner under twelve different ambient lighting conditions
Journal of Dentistry · 2023-07 · 134 · 104530
DOI: 10.1016/j.jdent.2023.104530 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Dentistry, 2023-07. Verfasst von Wenceslao Piedra-Cascón und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery and Medical-Surgery Specialitites, Doctoral Progamme in Dental Science, Stomatology Area, University of Santiago de Compostela,…. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Computer-Aided Design. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 46 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVES: The ambient lighting condition has been identified as an important factor that influences the accuracy of intraoral scanners (IOSs). The purpose of this study was to evaluate the influence of 12 different ambient lighting conditions on the accuracy of a confocal based IOS (PrimeScan). MATERIALS AND METHODS: A typodont was digitized using a laboratory scanner (L2i) to obtain a reference standard tessellation language (STLr) file. A restorative dentist recorded the scans using an IOS (PrimeScan) under 12 different ambient lighting conditions where the luminosity was measured using a light meter (LX1330B Light Meter). Twelve groups were created, namely 0-, 500-, 1000-, 2000-, 3000-, 4000-, 5000-, 6000-, 7000-, 8000-, 9000-, and 10 000 lux groups. Ten STL files were recorded per group. The STLr file was used as a reference with which to compare the distortion of the 120 STL files obtained using a software program (Meshlab). The normality Shapiro-Wilk test indicated that the distributions were not normal. Therefore, the nonparametric Kruskal-Wallis and pairwise multicomparison tests were used to analyze the data (α = 0.05). RESULTS: The group with the 1000 lux lighting condition obtained the smallest median ±interquartile range (IQR) with scanning distortion values of 69.5 ± 97.4 µm, followed by the 8000 lux group with a median ±IQR of 166.5 ± 318.1 µm. The 0 lx group presented the highest distortion values with a mean ±IQR of 355.5 ± 488.0 µm (p < 0.05). CONCLUSIONS: Ambient lighting conditions influenced the accuracy of the IOS tested. The highest accuracy values were obtained with 1000 lux. The lowest scanning accuracy was obtained with 0 lux.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery and Medical-Surgery Specialitites, Doctoral Progamme in Dental Science, Stomatology Area, University of Santiago de Compostela, Spain. Affiliate Faculty Esthetic Dentistry Program, Complutense Univ…
- Graduate Research Assistant J. Mike Walker '66 Department of Mechanical Engineering, Texas A&M University, College Station, TX, USA.
- Dental Materials Unit, Center for Dental and Oral Medicine, University of Zürich, Switzerland.
- Assistant Faculty J. Mike Walker '66 Department of Mechanical Engineering, Texas A&M University, College Station, TX, USA.
- Department of Restorative Dentistry, School of Dentistry, Graduate Prosthodontics, University of Washington, Seattle, Wash; Director of Research and Digital Dentistry, Kois Center, Seattle, Wash; and Adjunct Professor,…
- Department of Surgery and Medical-Surgery Specialities, Senior Lecturer in Planification and Management in Dental Clinics, Stomatology Area, Digital Dentistry Unit of the School of Dentistry, Faculty of Medicine and Den…
Zitieren
APA
Piedra-Cascón, W., Adhikari, R. R., Özcan, M., Krishnamurthy, V. R., Revilla-León, M., & Gallas-Torreira, M. (2023). Accuracy assessment (trueness and precision) of a confocal based intraoral scanner under twelve different ambient lighting conditions. Journal of Dentistry, 134, 104530. https://doi.org/10.1016/j.jdent.2023.104530
Vancouver
Piedra-Cascón W, Adhikari RR, Özcan M, Krishnamurthy VR, Revilla-León M, Gallas-Torreira M. Accuracy assessment (trueness and precision) of a confocal based intraoral scanner under twelve different ambient lighting conditions. Journal of Dentistry. 2023;134:104530. doi:10.1016/j.jdent.2023.104530
BibTeX
@article{PiedraCascn2023,
author = {Piedra-Cascón, Wenceslao and Adhikari, Riddhi R and Özcan, Mutlu and Krishnamurthy, Vinayak R and Revilla-León, Marta and Gallas-Torreira, Mercedes},