Enhancing Delivery Efficiency on the Magnetic Resonance-Linac: A Comprehensive Evaluation of Prostate Stereotactic Body Radiation Therapy Using Volumetric Modulated Arc Therapy
International Journal of Radiation Oncology*Biology*Physics · 2025
DOI: 10.1016/j.ijrobp.2024.10.028 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2025. Verfasst von Jeffrey E Snyder und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut; Department of Radiation Oncology, University of Iowa…. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Radiotherapy, Intensity-Modulated, Radiosurgery, Particle Accelerators. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 44 Literaturangaben. Gefördert von: Elekta, Nederlandse Organisatie voor Wetenschappelijk Onderzoek.
Abstract (Originalsprache)
PURPOSE: Long treatment sessions are a limitation within magnetic resonance imaging guided adaptive radiation therapy (MRIgART). This work aims for significantly enhancing the delivery efficiency on the magnetic resonance linear accelerator (MR-linac) by introducing dedicated optimization and delivery techniques for volumetric modulated arc therapy (VMAT). VMAT plan and delivery quality during MRIgART is compared with step-and-shoot intensity-modulated radiation therapy (IMRT) for prostate stereotactic body radiation therapy. METHODS AND MATERIALS: Ten patients with prostate cancer previously treated on a 1.5T MR-linac were retrospectively replanned to 36.25 Gy in 5 fractions using step-and-shoot IMRT and the clinical Hyperion optimizer within Monaco (Hyp-IMRT), the same optimizer with a VMAT technique (Hyp-VMAT), and a research-based optimizer called optimal fluence levels and pseudo gradient descent with VMAT (OFL+PGD-VMAT). The plans were then adapted onto each daily magnetic resonance imaging data set using 2 different optimization strategies to evaluate the adapt-to-position workflow: "optimize weights" (IMRT-Weights and VMAT-Weights) and "optimize shapes" (IMRT-Shapes and VMAT-Shapes). Treatment efficiency was evaluated by measuring optimization time, delivery time, and total time (optimization+delivery). Plan quality was assessed by evaluating organ at risk sparing. Ten patient plans were measured using a modified linac control system to assess delivery accuracy via a gamma analysis (2%/2 mm). Delivery efficiency was calculated as average dose rate divided by maximum dose rate. RESULTS: For Hyp-VMAT and OFL+PGD-VMAT, the total time was reduced by 124 ± 140 seconds (P = .020) and 459 ± 110 seconds (P < .001), respectively, as compared with the clinical Hyp-IMRT group. Speed enhancements were also measured for adapt-to-position with reductions in total time of 404 ± 55 (P < .001) for VMAT-Weights as compared with the clinical IMRT-Shapes group. Bladder and rectum dosimetric volume histogram (DVH) points were within 1.3% or 0.8 cc for each group. All VMAT plans had gamma passing rates greater than 96%. The delivery efficiency of VMAT plans was 89.7 ± 2.7 % compared with 50.0 ± 2.2 % for clinical IMRT. CONCLUSIONS: Incorporating VMAT into MRIgART will significantly reduce treatment session times while maintaining equivalent plan quality.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut; Department of Radiation Oncology, University of Iowa, Iowa City, Iowa. Electronic address: jeffrey.snyder@yale.edu.
- Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands.
- Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands; Elekta AB, Stockholm, Sweden.
- Department of Radiation Oncology, University of Iowa, Iowa City, Iowa.
Förderung
- Elekta
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Zitieren
APA
Snyder, J. E., Fast, M. F., Uijtewaal, P., Borman, P. T. S., Woodhead, P., St-Aubin, J., Smith, B., Shepard, A., Raaymakers, B. W., & Hyer, D. E. (2025). Enhancing Delivery Efficiency on the Magnetic Resonance-Linac: A Comprehensive Evaluation of Prostate Stereotactic Body Radiation Therapy Using Volumetric Modulated Arc Therapy. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2024.10.028
Vancouver
Snyder JE, Fast MF, Uijtewaal P, Borman PTS, Woodhead P, St-Aubin J, et al. Enhancing Delivery Efficiency on the Magnetic Resonance-Linac: A Comprehensive Evaluation of Prostate Stereotactic Body Radiation Therapy Using Volumetric Modulated Arc Therapy. International Journal of Radiation Oncology*Biology*Physics. 2025. doi:10.1016/j.ijrobp.2024.10.028
BibTeX
@article{Snyder2025,
author = {Snyder, Jeffrey E and Fast, Martin F and Uijtewaal, Prescilla and Borman, Pim T S and Woodhead, Peter and St-Aubin, Joël and Smith, Blake and Shepard, Andrew and Raaymakers, Bas W and Hyer, Daniel E},
title = {Enhancing Delivery Efficiency on the Magnetic Resonance-Linac: A Comprehensive Evaluation of Prostate Stereotactic Body Radiation Therapy Using Volumetric Modulated Arc Therapy},
journal = {International Journal of Radiation Oncology*Biology*Physics},
year = {2025},
doi = {10.1016/j.ijrobp.2024.10.028}
}
EUVIMED