Strategies for enhancing delivery efficiency on MR‐Linac: A dosimetric study and historical plan review
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70638 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Ying Zhang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Medical Artificial Intelligence and Automation (MAIA) Lab & Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Prostatic Neoplasms, Radiotherapy, Intensity-Modulated, Liver Neoplasms. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The Elekta Unity MR‐Linac enables online adaptive radiotherapy (oART) with integrated 1.5T MRI. Given its relatively low dose rate, identifying optimal IMRT planning parameters is essential to achieve efficient delivery. This study investigates parameter optimization strategies that enhance delivery efficiency while maintaining high dosimetric quality.
Methods
A dosimetric study was designed using ten prostate simultaneous integrated boost (SIB) patients (45 Gy/40 Gy in five fractions) and 10 liver SBRT patients (42–54 Gy in 3–5 fractions). Each patient has three plans: (1) Clinical, (2) Refa (segment‐limited, ≤120 segments), and (3) Refb (MU‐modulated plan with larger minimum segment width (0.7 cm or 1 cm) and higher fluence map smoothing level (medium or high)). A linear equation for estimating delivery time was established and validated against 1173 retrospective Unity plans based on total segments and Monitor Units (MUs). All the study plans were evaluated by comparing delivery efficiency related metrics (total number of segments, MUs, estimated delivery time), and plan quality using dose–volume histogram (DVH) metrics for targets and organs‐at‐risk (OARs). Additionally, an online adaptive study was performed on representative cases: the three reference plans were adapted to a daily image, and the resulting adaptive plans were assessed for optimization time and plan quality against clinical criteria.
Results
Using the derived delivery time estimation equation on 173 unseen test cases, the mean absolute error (MAE) between estimated and recorded times was 0.9 min, with 84% of cases with a prediction error within ±1 min, and with 95% of cases falling within ±2 min. Both strategies reduced delivery complexity compared with clinical plans. For prostate cases, Refa and Refb lowered segment counts to
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Artificial Intelligence and Automation (MAIA) Lab & Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA.
Zitieren
APA
Zhang, Y., Tang, S., Kabat, C., Visak, J., Chiu, T., Li, R., Jiang, S., & Lin, M. H. (2026). Strategies for enhancing delivery efficiency on MR‐Linac: A dosimetric study and historical plan review. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70638
Vancouver
Zhang Y, Tang S, Kabat C, Visak J, Chiu T, Li R, et al. Strategies for enhancing delivery efficiency on MR‐Linac: A dosimetric study and historical plan review. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70638
BibTeX
@article{Zhang2026,
author = {Zhang, Ying and Tang, Shanshan and Kabat, Christopher and Visak, Justin and Chiu, Tsuicheng and Li, Ruiqi and Jiang, Steve and Lin, Mu-Han},
title = {Strategies for enhancing delivery efficiency on MR‐Linac: A dosimetric study and historical plan review},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70638}
}
EUVIMED