EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dose-dependent ventricular lesion formation using a novel large-area pulsed field ablation catheter: A preclinical feasibility study

Thomas Kueffer, Daniela Casoni, Christine Goepfert, Olgica Beslac, Chiara Parodi, David Ramirez, Kara Garrott, Brendan Koop, Sean Coe, Nathan Hagstrom, Guillaume Gibert, Laurent Roten und 2 weitere

Heart Rhythm · 2025-09 · 22 (9) · 2322-2330

DOI: 10.1016/j.hrthm.2025.02.017 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm, 2025-09. Verfasst von Thomas Kueffer und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland; SITEM Center for Translational Medicine and B…. Hauptthemen laut MEDLINE: Catheter Ablation, Heart Ventricles. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Pulsed field ablation (PFA) has shown promising data in terms of safety and procedural efficiency for pulmonary vein isolation. Large-area focal PFA catheter designs might be suitable to deliver deep and durable lesions in ventricular myocardium. OBJECTIVE: We aimed to investigate the dose-response of a novel large-area focal 3-dimensional (3D)-enabled map-and-ablate PFA catheter for ventricular ablation in a chronic preclinical swine model. METHODS: An 8F catheter with a 9-mm hexaspline tip was used for 3D mapping of both ventricles in a porcine model. Using a PFA generator with a proprietary waveform optimized for the catheter, left and right ventricular lesions were placed with either a monopolar or bipolar ablation vector and with 1, 2, or 4 applications per site (2.0 kV/application). Tissue contact was ensured by intracardiac echocardiography and electrograms. The animals were kept alive for 1 week. Ablation lesions were assessed macroscopically after triphenyl tetrazolium chloride staining and by histopathology. RESULTS: A total of 69 chronic ventricular lesions from 7 pigs were available for analysis. By stacking 4 PFA applications rather than a single application, median chronic lesion depth increased from 4.8 mm (interquartile range [IQR], 4.1-5.6 mm) to 5.5 mm (IQR, 5.0-6.2 mm; P = .06) with bipolar ablation and from 4.9 mm (IQR, 4.4-5.2 mm) to 6.5 mm (IQR, 5.9-6.9 mm; P = .002) with monopolar ablation. On histologic evaluation, lesion borders were clearly demarcated, with vessels and nerves preserved. CONCLUSION: A novel large-area focal ablation catheter with the ability for 3D mapping and PFA was able to create dose-dependent deep ventricular lesions durable 1 week after ablation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheter AblationHeart VentriclesAnimalsSwineFeasibility StudiesDisease Models, AnimalEquipment Design

Stichwörter der Autor:innen

Bipolar PFACatheter ablationLesion depthMap and ablateMonopolar PFAPulsed field ablationVentricular ablation

Autor:innen und Institutionen

  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland; SITEM Center for Translational Medicine and Biomedical Entrepreneurship, University of Bern, Bern, Switzerland. Ele…
  • Experimental Surgical Facility, Experimental Animal Center, University of Bern, Bern, Switzerland.
  • COMPATH, Institute of Animal Pathology, University of Bern, Bern, Switzerland.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland; SITEM Center for Translational Medicine and Biomedical Entrepreneurship, University of Bern, Bern, Switzerland; Exp…
  • Boston Scientific Corporation, St Paul, Minnesota.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

Zitieren

APA

Kueffer, T., Casoni, D., Goepfert, C., Beslac, O., Parodi, C., Ramirez, D., Garrott, K., Koop, B., Coe, S., Hagstrom, N., Gibert, G., Roten, L., Haeberlin, A., & Reichlin, T. (2025). Dose-dependent ventricular lesion formation using a novel large-area pulsed field ablation catheter: A preclinical feasibility study. Heart Rhythm, 22(9), 2322-2330. https://doi.org/10.1016/j.hrthm.2025.02.017

Vancouver

Kueffer T, Casoni D, Goepfert C, Beslac O, Parodi C, Ramirez D, et al. Dose-dependent ventricular lesion formation using a novel large-area pulsed field ablation catheter: A preclinical feasibility study. Heart Rhythm. 2025;22(9):2322-2330. doi:10.1016/j.hrthm.2025.02.017

BibTeX
@article{Kueffer2025,
  author = {Kueffer, Thomas and Casoni, Daniela and Goepfert, Christine and Beslac, Olgica and Parodi, Chiara and Ramirez, David and Garrott, Kara and Koop, Brendan and Coe, Sean and Hagstrom, Nathan and Gibert, Guillaume and Roten, Laurent and Haeberlin, Andreas and Reichlin, Tobias},
  title = {Dose-dependent ventricular lesion formation using a novel large-area pulsed field ablation catheter: A preclinical feasibility study},
  journal = {Heart Rhythm},
  year = {2025},
  volume = {22},
  number = {9},
  pages = {2322--2330},
  doi = {10.1016/j.hrthm.2025.02.017}
}
Kostenlos starten →