Bipolar endo-epicardial RF ablation: Animal feasibility study
Heart Rhythm · 2024
DOI: 10.1016/j.hrthm.2024.02.009 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart Rhythm, 2024. Verfasst von Paweł Derejko und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Medicover Hospital, Warsaw, Poland; Department of Cardiac Arrhythmias, National Institute of Cardiology, Warsaw, Poland. El…. Hauptthemen laut MEDLINE: Feasibility Studies, Pericardium, Endocardium, Disease Models, Animal. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 14 Literaturangaben. Gefördert von: Biosense Webster.
Abstract (Originalsprache)
BACKGROUND: Bipolar radiofrequency ablation (B-RFA) is a method used to treat the arrhythmia substrate resistant to unipolar ablation. Few studies have addressed endo-epicardial B-RFA. OBJECTIVE: The aim of the study was to evaluate chronic lesions resulting from endo-epicardial B-RFA and to determine optimal settings for such procedures in an animal model. METHODS: In 7 pigs, up to 5 radiofrequency applications per animal were performed with 2 electrodes placed on both sides of the left ventricular free wall. Current was delivered for 60 seconds by a generator dedicated for B-RFA with power settings of 25, 30, 35, 40, and 50 W. RESULTS: At 12 weeks after ablation, 31 lesions were assessed. Their maximal cross-sectional area ranged from 7.2 to 68 mm2 and correlated with total power delivered (r = 0.53), with temperature increment at the endocardial catheter (r = 0.65), and inversely with temperature decrement at the epicardial catheter (r = 0.54). For power values between 30 and 40 W, the lesion area did not differ significantly (P = .92). Lesion depth ranged from 1.9 to 11 mm and correlated with impedance decrement (r = 0.5). Lesions were transmural in 8 cases. Lesion depth/wall thickness ratio was on average 0.6 ± 0.3, with the smallest value for 25 W (0.5 ± 0.3) and the largest for 50 W (0.8 ± 0.3). Steam pops occurred at a power range of 30-50 W, with an incidence of 1 in 5 applications, with 1 case of fatal tamponade at 40 W. Impedance decrement, endocardial catheter temperature increment, and endocardial electrogram amplitude decrement were greater during applications with steam pops. CONCLUSION: Chronic lesions resulting from endo-epicardial B-RFA appear smaller and less often transmural compared with acute lesions described in the literature. The incidence of steam pops during endo-epicardial B-RFA is relatively high even at low powers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Medicover Hospital, Warsaw, Poland; Department of Cardiac Arrhythmias, National Institute of Cardiology, Warsaw, Poland. Electronic address: pderejko@yahoo.com.
- Faculty of Veterinary Medicine, University of Environmental and Life Sciences, Wroclaw, Poland.
- Department of Cardiology, Centre for Postgraduate Medical Education, Grochowski Hospital, Warsaw, Poland.
- Department of Cardiology, Medicover Hospital, Warsaw, Poland.
- Biosense Webster, Warsaw, Poland.
Förderung
- Biosense Webster
Zitieren
APA
Derejko, P., Janus, I., Kułakowski, P., Kuśnierz, J., Baran, J., Hangiel, U., Proszek, J., Frydrychowski, P., Michałek, M., & Noszczyk-Nowak, A. (2024). Bipolar endo-epicardial RF ablation: Animal feasibility study. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2024.02.009
Vancouver
Derejko P, Janus I, Kułakowski P, Kuśnierz J, Baran J, Hangiel U, et al. Bipolar endo-epicardial RF ablation: Animal feasibility study. Heart Rhythm. 2024. doi:10.1016/j.hrthm.2024.02.009
BibTeX
@article{Derejko2024,
author = {Derejko, Paweł and Janus, Izabela and Kułakowski, Piotr and Kuśnierz, Jacek and Baran, Jakub and Hangiel, Urszula and Proszek, Joanna and Frydrychowski, Piotr and Michałek, Marcin and Noszczyk-Nowak, Agnieszka},
title = {Bipolar endo-epicardial RF ablation: Animal feasibility study},
journal = {Heart Rhythm},
year = {2024},
doi = {10.1016/j.hrthm.2024.02.009}
}
EUVIMED