Temperature-based rate response in a leadless pacemaker system
Heart Rhythm · 2025
DOI: 10.1016/j.hrthm.2024.11.032 ↗Kurz gefasst
Klinische Studie (Phase II) in der Zeitschrift Heart Rhythm, 2025. Verfasst von Mayer Y Rashtian und 13 weiteren Autor:innen. Beteiligte Einrichtung: Huntington Memorial Hospital, Pasadena, California. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Heart Rate, Pacemaker, Artificial. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Abbott Laboratories Inc.
Abstract (Originalsprache)
BACKGROUND: A new dual-chamber leadless pacemaker (DR-LP) system, composed of 2 implantable devices in the right ventricle and right atrium, uses a less common temperature-based rate-response sensor. There is a need to understand the effectiveness of the rate response during exercise in both the ventricular (VR-LP) and atrial (AR-LP) devices. OBJECTIVE: We sought to determine whether temperature-based rate-responsive pacing is proportional to metabolic workload during an exercise test in a leadless pacemaker system. METHODS: After 6 weeks of implantation, we administered a treadmill exercise protocol to eligible participants concurrently enrolled in the LEADLESS II-Phase 2 and Aveir DR i2i studies. Programmed settings were optimized after a prior 6-minute walk test. We evaluated the ventricular and atrial rate-response sensors in participants implanted with the VR-LP and DR-LP system, respectively. For each device, the normalized slopes of sensor-indicated rate vs metabolic workload were aggregated across all analyzable patients. If the mean slope's 95% confidence interval (CI) fell within the prespecified 0.65 and 1.35 acceptance range, the rate response was considered proportional to metabolic demand. RESULTS: Seventeen participants had a mean ventricular rate-response slope of 0.93 ± 0.29 (CI, 0.78-1.08), which fell within the acceptance criteria (P = .001). Twenty participants had a mean atrial rate-response slope of 0.91 ± 0.28 (CI, 0.78-1.05), also falling within the prespecified criteria (P < .001). CONCLUSION: The temperature-based sensor in a dual-chamber leadless pacemaker system was shown to be effective at modulating pacing rate in response to increased metabolic demand for right ventricular and atrial devices. GOV IDENTIFIER: NCT04559945 (LEADLESS II-Phase 2 study) and NCT05252702 (Aveir DR i2i study).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Huntington Memorial Hospital, Pasadena, California.
- Weill Cornell Medicine/NewYork-Presbyterian Hospital, New York, New York. Electronic address: jei9008@med.cornell.edu.
- Foothills Medical Centre, Calgary, Alberta, Canada.
- Mount Sinai Fuster Heart Hospital, Icahn School of Medicine at Mount Sinai, New York, New York.
- HonorHealth Cardiac Arrhythmia Group, Scottsdale, Arizona.
- CHRU Albert Michallon, Grenoble, France.
Förderung
- Abbott Laboratories Inc
Zitieren
APA
Rashtian, M. Y., Ip, J. E., Exner, D. V., Reddy, V. Y., Doshi, R. N., Defaye, P., Canby, R. C., Bongiorni, M. G., Shoda, M., Hindricks, G., Nevo, J. R., Ligon, D., Bulusu, A., & Knops, R. E. (2025). Temperature-based rate response in a leadless pacemaker system. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2024.11.032
Vancouver
Rashtian MY, Ip JE, Exner DV, Reddy VY, Doshi RN, Defaye P, et al. Temperature-based rate response in a leadless pacemaker system. Heart Rhythm. 2025. doi:10.1016/j.hrthm.2024.11.032
BibTeX
@article{Rashtian2025,
author = {Rashtian, Mayer Y and Ip, James E and Exner, Derek V and Reddy, Vivek Y and Doshi, Rahul N and Defaye, Pascal and Canby, Robert C and Bongiorni, Maria Grazia and Shoda, Morio and Hindricks, Gerhard and Nevo, Jordan R and Ligon, David and Bulusu, Anu and Knops, Reinoud E},
title = {Temperature-based rate response in a leadless pacemaker system},
journal = {Heart Rhythm},
year = {2025},
doi = {10.1016/j.hrthm.2024.11.032}
}
EUVIMED