Beam data commissioning, validation and clinical implementation of SRS gyroscopic system

Catarina Souto, Borja Aguilar, Vânia Dias, Enrique Pascual, Gabriel Farinha, Carlos Marcelino, Carlos Rodrigues, Pedro Teles, Alberto Perez-Rozos, Morena Sallabanda, Kita Sallabanda

Physica Medica · 2025

DOI: 10.1016/j.ejmp.2025.104952 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Physica Medica, 2025. Verfasst von Catarina Souto und 10 weiteren Autor:innen. Beteiligte Einrichtung: IRCA, Hospital Viamed Santa Elena, Madrid, Spain; Mercurius Health, Porto Salvo, Portugal; DFA, Faculdade de Ciências da Universidade do Porto, Porto…. Hauptthemen laut MEDLINE: Radiosurgery. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Fundação para a Ciência e a Tecnologia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The very unconventional design and features of the new gyroscopic radiosurgery ZAP-X® system (ZAP Surgical Systems) generates a challenge for beam data acquisition, characterization and validation. Treatment Planning System commissioning was performed with different detectors, filling the absence of methodologies and correction factors published in international guidelines. The congruence of detectors response was compared with corresponding SciMoCaTM Monte Carlo (MC) synthetic data (Scientific RT GmbH). A machine and patient-specific quality assurance (QA) protocol was also implemented. MATERIALS AND METHODS: All data were acquired using a microsilicon and a microdiamond detectors, three ionization chambers (PTW Freiburg) and a plastic scintillation detector (PSD) (Blue Physics LLC). For patient-specific QA, an anthropomorphic phantom (RT Safe Inc), a 2D array (Sun Nuclear Corporation) and SciMoCaTM were used for 1D dose verification, 2D and 3D evaluation gamma assessment, respectively. RESULTS: Agreement between detectors, for all PDDs, was better than 1% beyond the depth of dmax. For profiles, the microsilicon detector showed steeper penumbras due to the smaller detector width. A good congruence with MC data was found for the solid-state detectors. Output factors measurements demonstrated strong agreement between SciMoCaTM data and PSD data, with a maximum difference of 2.1%. The patient QA results validated the quality of the commissioning data. CONCLUSIONS: Our results showed strong agreement between different measurement methods. MC simulations proved to be an important tool for commissioning validation. The methods and materials discussed in this article may be of use to other facilities where this SRS gyroscope system is installed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryHumansMonte Carlo MethodPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy DosageQuality ControlRadiometry

Stichwörter der Autor:innen

CommissioningSRSSmall field dosimetryZAP-X®

Autor:innen und Institutionen

  • IRCA, Hospital Viamed Santa Elena, Madrid, Spain; Mercurius Health, Porto Salvo, Portugal; DFA, Faculdade de Ciências da Universidade do Porto, Porto, Portugal. Electronic address: anacatarinasouto@gmail.com.
  • IRCA, Hospital Viamed Santa Elena, Madrid, Spain; Mercurius Health, Porto Salvo, Portugal.
  • IRCA, Hospital Viamed Santa Elena, Madrid, Spain.
  • Mercurius Health, Porto Salvo, Portugal.
  • DFA, Faculdade de Ciências da Universidade do Porto, Porto, Portugal; CI-IPO Porto, Porto, Portugal.
  • IRCA, Hospital Viamed Santa Elena, Madrid, Spain; Mercurius Health, Porto Salvo, Portugal; Servicio Andaluz de Salud, Sevilla, Spain.

Förderung

  • Fundação para a Ciência e a Tecnologia

Zitieren

APA

Souto, C., Aguilar, B., Dias, V., Pascual, E., Farinha, G., Marcelino, C., Rodrigues, C., Teles, P., Perez-Rozos, A., Sallabanda, M., & Sallabanda, K. (2025). Beam data commissioning, validation and clinical implementation of SRS gyroscopic system. Physica Medica. https://doi.org/10.1016/j.ejmp.2025.104952

Vancouver

Souto C, Aguilar B, Dias V, Pascual E, Farinha G, Marcelino C, et al. Beam data commissioning, validation and clinical implementation of SRS gyroscopic system. Physica Medica. 2025. doi:10.1016/j.ejmp.2025.104952

BibTeX
@article{Souto2025,
  author = {Souto, Catarina and Aguilar, Borja and Dias, Vânia and Pascual, Enrique and Farinha, Gabriel and Marcelino, Carlos and Rodrigues, Carlos and Teles, Pedro and Perez-Rozos, Alberto and Sallabanda, Morena and Sallabanda, Kita},
  title = {Beam data commissioning, validation and clinical implementation of SRS gyroscopic system},
  journal = {Physica Medica},
  year = {2025},
  doi = {10.1016/j.ejmp.2025.104952}
}
Kostenlos starten →