Experimental and Monte Carlo based dosimetric investigation of a novel 3 mm radiosurgery 3 MV beam using the microSilicon detector
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14388 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Katrin Saße und 10 weiteren Autor:innen. Beteiligte Einrichtung: University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiosurgery, Particle Accelerators, Radiotherapy Dosage. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The ZAP‐X system is a novel gyroscopic radiosurgical system based on a 3 MV linear accelerator and collimator cones with a diameter between 4 and 25 mm. Advances in imaging modalities to detect small and early‐stage pathologies allow for an early and less invasive treatment, where a smaller collimator matching the anatomical target could provide better sparing of surrounding healthy tissue.
Purpose
A novel 3 mm collimator cone for the ZAP‐X was developed. This study aims to investigate the usability of a commercial diode detector (microSilicon) for the dosimetric characterization of this small collimator cone; and to investigate the underlying small field perturbation effects.
Methods
Profile measurements in five depths as well as PDD and output ratio measurements were performed with a microSilicon detector and radiochromic EBT3 films. In addition, comprehensive Monte Carlo simulations were performed to validate the measurement observations and to quantify the perturbation effects of the microSilicon detector in these extremely small field conditions.
Results
It is shown that the microSilicon detector enables an accurate dosimetric characterization of the 3 mm beam. The profile parameters, such as the FWHM and 20%–80% penumbra width, agree within 0.1 to 0.2 mm between film and detector measurements. The output ratios agree within the measurement uncertainty between microSilicon detector and films, whereas the comparisons of the PDD results show good agreement with the Monte Carlo simulations. The analysis of the perturbation factors of the microSilicon detector reveals a small field correction factor of approximately 3% for the 3 mm circular beam and a correction factor smaller than 1.5% for field diameters above 3 mm.
Conclusions
It could be shown that the microSilicon detector is well‐suitable for the characterization of the new 3 mm circular beam of the ZAP‐X system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
- Bonifatius Hospital Lingen, Lingen, Germany.
- Department of Neurosurgery, Stanford University School of Medicine, Palo Alto, California, USA.
- ZAP Surgical Systems, San Carlos, California, USA.
- PTW-Freiburg, Freiburg, Germany.
Zitieren
APA
Saße, K., Albers, K., Klassen, P. D., Marianyagam, N. J., Weidlich, G., Schneider, M. B., Chang, S., Adler, J., Poppe, B., Looe, H. K., & Eulenstein, D. (2024). Experimental and Monte Carlo based dosimetric investigation of a novel 3 mm radiosurgery 3 MV beam using the microSilicon detector. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14388
Vancouver
Saße K, Albers K, Klassen PD, Marianyagam NJ, Weidlich G, Schneider MB, et al. Experimental and Monte Carlo based dosimetric investigation of a novel 3 mm radiosurgery 3 MV beam using the microSilicon detector. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14388
BibTeX
@article{Sae2024,
author = {Saße, Katrin and Albers, Karina and Klassen, Peter Douglas and Marianyagam, Neelan J and Weidlich, Georg and Schneider, M Bret and Chang, Steven and Adler, John and Poppe, Björn and Looe, Hui Khee and Eulenstein, Daniela},
title = {Experimental and Monte Carlo based dosimetric investigation of a novel 3 mm radiosurgery 3 MV beam using the microSilicon detector},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14388}
}
EUVIMED