Determination of the beam quality correction factor kQmsr${k}_{{Q}_{{\mathrm{msr}}}}$ for the PTW Semiflex 3D ionization chamber for the reference dosimetry at ZAP‐X
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14610 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Katrin Saße und 5 weiteren Autor:innen. Beteiligte Einrichtung: University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany. Hauptthemen laut MEDLINE: Radiometry, Particle Accelerators, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The self‐shielding radiosurgery system ZAP‐X consists of a 3 MV linear accelerator and eight round collimators. For this system, it is a common practice to perform the reference dosimetry using the largest 25 mm diameter collimator at a source‐to‐axis distance (SAD) of 45 cm with the PTW Semiflex3D chamber placed at a measurement depth of 7 mm in water. Existing dosimetry protocols do not provide correction for these measurement conditions. Therefore, Monte Carlo simulations were performed to quantify the associated beam quality correction factor .
Methods
The of the Semiflex3D chamber was computed from the ratio of the absorbed doses in a water voxel and in the sensitive air volume of the chamber simulated using a 60 Co spectrum as the calibration beam quality ( Q ref ) and the spectrum of the ZAP‐X 3 MV photon beam ( Q msr ). was computed as a function of measurement depth from 4 to 50 mm. Furthermore, detailed simulations were performed to determine the individual chamber's perturbation correction factors by modifying the chamber's model step‐wise.
Results
All perturbation correction factors, except S w,air ⋅ P fl , show depth‐dependent behavior up to a depth of 15 mm. In particular, the volume‐averaging P vol and density P dens perturbation correction factors and, consequently, the resulting gradient perturbation correction factor P gr = P vol ∙ P dens increase with decreasing measurement depth. Therefore, is larger than unity, amounting to at 7 mm measurement depth. At larger depths (> 15 mm), the can be considered as constant.
Conclusion
At small measurement depths, was found to be depth‐dependent with values larger than unity due to the gradient‐related perturbation factors. Therefore, the uncertainty related to the chamber's positioning can be reduced by performing the reference dosimetry at ZAP‐X at depths larger than 15 mm, where can be regarded as depth independent.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
- Zap Surgical, Medical Physics Department, PTW-Freiburg, Freiburg, Germany.
- ZAP Surgical, San Carlos, California, USA.
Zitieren
APA
Saße, K., Albers, K., Eulenstein, D., Weidlich, G., Poppe, B., & Looe, H. K. (2025). Determination of the beam quality correction factor kQmsr${k}_{{Q}_{{\mathrm{msr}}}}$ for the PTW Semiflex 3D ionization chamber for the reference dosimetry at ZAP‐X. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14610
Vancouver
Saße K, Albers K, Eulenstein D, Weidlich G, Poppe B, Looe HK. Determination of the beam quality correction factor kQmsr${k}_{{Q}_{{\mathrm{msr}}}}$ for the PTW Semiflex 3D ionization chamber for the reference dosimetry at ZAP‐X. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14610
BibTeX
@article{Sae2025,
author = {Saße, Katrin and Albers, Karina and Eulenstein, Daniela and Weidlich, Georg and Poppe, Björn and Looe, Hui Khee},
title = {Determination of the beam quality correction factor kQmsr$k_Q_\mathrmmsr$ for the PTW Semiflex 3D ionization chamber for the reference dosimetry at ZAP‐X},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14610}
}
EUVIMED