EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Accuracy of intraocular lens formulas using a sum-of-segments axial length biometer in an Asian population

John Tsia-Chuen Kan, Stephen Stewart, John Xin Hao Wong, Tun Kuan Yeo

Eye Discovery · 2026

DOI: 10.1016/j.edisc.2026.100025 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Eye Discovery, 2026. Verfasst von John Tsia-Chuen Kan und 3 weiteren Autor:innen. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare biometric measurements obtained using two optical biometers, the Argos and the IOLMaster 700, and to evaluate the accuracy of intraocular lens (IOL) power calculation formulas based on Argos measurements in an Asian population.

Methods

Pre-operative biometric data acquired using both the Argos and IOLMaster 700 biometers in patients undergoing cataract surgery were analyzed. Post-operative prediction error was calculated by subtracting the predicted refraction generated by each IOL formula, based on Argos-derived biometric data, from the actual manifest refraction measured 1 month after surgery. Both traditional and modern IOL formulas were evaluated, including formulas specifically designed for sum-of-segment axial length (AL) measurements, such as the Barrett True Axial Length and EVO 2.0 formulas. Subgroup analyses were performed for short eyes with AL 26 mm.

Results

A total of 215 eyes from 215 patients were included. Compared with measurements obtained using the IOLMaster 700, AL and white-to-white distance measured by the Argos were lower, whereas anterior chamber depth and mean keratometry values were significantly higher. AL measurements differed significantly between the two biometers in both short and long eyes. Most IOL power calculation formulas demonstrated good performance when using Argos measurements, including in eyes with short and long ALs. On linear regression analysis, the EVO and Holladay I formulas showed no statistically significant AL-dependent bias in prediction error.

Conclusion

IOL power calculation formulas demonstrate good performance when applied to sum-of-segment AL measurements obtained using the Argos biometer. Overall, formulas specifically designed for sum-of-segment AL measurements yielded the best predictive accuracy.

Zitieren

APA

Kan, J. T., Stewart, S., Wong, J. X. H., & Yeo, T. K. (2026). Accuracy of intraocular lens formulas using a sum-of-segments axial length biometer in an Asian population. Eye Discovery. https://doi.org/10.1016/j.edisc.2026.100025

Vancouver

Kan JT, Stewart S, Wong JXH, Yeo TK. Accuracy of intraocular lens formulas using a sum-of-segments axial length biometer in an Asian population. Eye Discovery. 2026. doi:10.1016/j.edisc.2026.100025

BibTeX
@article{Kan2026,
  author = {Kan, John Tsia-Chuen and Stewart, Stephen and Wong, John Xin Hao and Yeo, Tun Kuan},
  title = {Accuracy of intraocular lens formulas using a sum-of-segments axial length biometer in an Asian population},
  journal = {Eye Discovery},
  year = {2026},
  doi = {10.1016/j.edisc.2026.100025}
}
Kostenlos starten →