The manufacturing method and guidance system affects the accuracy of implant placement by inexperienced clinicians: a comparative study

Hannah Dhaliwal, Alyssa Margolian, Selina Tran, Andrew B. Cameron, Lavanya Ajay Sharma, Ajay Sharma

Digital Dentistry Journal · 2025

DOI: 10.1016/j.ddj.2025.100014 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Digital Dentistry Journal, 2025. Verfasst von Hannah Dhaliwal und 5 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluates the accuracy of dental implant placements performed by inexperienced clinicians with surgical guides manufactured at different orientation (0°, 45°, and 90°) using three systems: Fully-guided (FG), pilot-guided (PG), and freehand (FH). The research aims to determine whether these clinicians can achieve precise implant placements in artificial mandibles Three dentistry students with four years of general clinical experience placed a total of 270 implants in 45 modified mandibular models. Each student placed 30 implants using FG, PG, and FH methods with surgical guide manufactured at 0°, 45°, and 90° angulations. Virtual implant planning was conducted using 3Shape software, and accuracy was evaluated by comparing post-placement scans to baseline plans. Statistical analysis included two-way ANOVA to assess the impact of placement method and printing angulation on the implant deviation angle, followed by Tukey post hoc testing for significant differences. The findings indicate that the FG system yields the highest accuracy with the minimum angle error of 0.12 degrees, followed by the PG and FH systems. Additionally, 3D-printed surgical guides at a 0° angulation demonstrated superior placement accuracy compared to those at 45° and 90°. These results have significant implications for clinical practice and training, suggesting that FG systems are preferable for novice clinicians and that optimal printing orientation enhances implant placement accuracy.

Zitieren

APA

Dhaliwal, H., Margolian, A., Tran, S., Cameron, A. B., Sharma, L. A., & Sharma, A. (2025). The manufacturing method and guidance system affects the accuracy of implant placement by inexperienced clinicians: a comparative study. Digital Dentistry Journal. https://doi.org/10.1016/j.ddj.2025.100014

Vancouver

Dhaliwal H, Margolian A, Tran S, Cameron AB, Sharma LA, Sharma A. The manufacturing method and guidance system affects the accuracy of implant placement by inexperienced clinicians: a comparative study. Digital Dentistry Journal. 2025. doi:10.1016/j.ddj.2025.100014

BibTeX
@article{Dhaliwal2025,
  author = {Dhaliwal, Hannah and Margolian, Alyssa and Tran, Selina and Cameron, Andrew B. and Sharma, Lavanya Ajay and Sharma, Ajay},
  title = {The manufacturing method and guidance system affects the accuracy of implant placement by inexperienced clinicians: a comparative study},
  journal = {Digital Dentistry Journal},
  year = {2025},
  doi = {10.1016/j.ddj.2025.100014}
}
Kostenlos starten →