Exploring strategies to rapidly identify false positives in high-sensitivity cardiac troponin I assay: A prospective study
Clinica Chimica Acta · 2025-01 · 565 · 119996
DOI: 10.1016/j.cca.2024.119996 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinica Chimica Acta, 2025-01. Verfasst von Anthony Desaegher und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biochemistry, Bordeaux University Hospital, Place Amélie Raba Léon, 33076 Bordeaux, France. Hauptthemen laut MEDLINE: Troponin I. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Cardiac troponin is the pivotal biomarker of myocardial injury, playing a central role in the diagnosis of acute coronary syndrome and various clinical situations. Nevertheless, challenges arise when patients exhibit elevated cardiac troponin levels in the absence of evident cardiac origin, as evidenced by extensive cardiac exploration, which suggests the presence of an interfering factor. Despite the high performance of high-sensitive cardiac troponin immunoassays, these tests remain susceptible to interferences that may lead to false-positives. METHODS: In the period between July 2021 and July 2024, 8129 patients were hospitalized in the cardiology departments of Bordeaux University Hospital with positive results for high-sensitivity cardiac troponin I. Among them, 15 patients were suspected of having false-positive results, based on clinical and biological observations. RESULTS: In this subpopulation, we evaluated prospectively various techniques, including serial dilutions, antibody-binding tubes, and alternative immunoassays (for cardiac troponin I and T) with the objective of identifying any potential analytical interference in their high-sensitive cardiac troponin I measurements. Our investigations revealed that 12 out of 15 suspected cases exhibited an interference on the high-sensitive cardiac troponin I assay. CONCLUSION: In conclusion, we propose an original algorithm designed to identify high-sensitive cardiac troponin I false-positives. This algorithm can help clinicians to make prompt and informed decisions about patient care, and to avoid erroneous clinical interventions that may result from such interferences.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biochemistry, Bordeaux University Hospital, Place Amélie Raba Léon, 33076 Bordeaux, France.
- Department of Biochemistry, Bordeaux University Hospital, Place Amélie Raba Léon, 33076 Bordeaux, France; Univ. Bordeaux, CNRS, CRMSB, UMR 5536, F-33000 Bordeaux, France.
- Cellule de Biologie Délocalisée, Bordeaux University Hospital, Place Amélie Raba Léon, 33076 Bordeaux, France.
- Department of Biochemistry, Bergonié Cancer Institute, F-33000 Bordeaux, France.
- Department of Cardiology, Bordeaux University Hospital, 1 rue Jean Burguet, 33000 Bordeaux, France.
- Department of Biochemistry, Bordeaux University Hospital, Place Amélie Raba Léon, 33076 Bordeaux, France; Univ. Bordeaux, INSERM, BRIC, U1312, F-33000 Bordeaux, France.
Zitieren
APA
Desaegher, A., Marin, V., Beauvieux, M. C., Colombiès, B., Lauga, M., Alloug, S., Kalkan, S., Castaing-Mouhica, G., Lacape, G., Rucheton, B., Doublet, J., Dabernat, S., & Bats, M. L. (2025). Exploring strategies to rapidly identify false positives in high-sensitivity cardiac troponin I assay: A prospective study. Clinica Chimica Acta, 565, 119996. https://doi.org/10.1016/j.cca.2024.119996
Vancouver
Desaegher A, Marin V, Beauvieux MC, Colombiès B, Lauga M, Alloug S, et al. Exploring strategies to rapidly identify false positives in high-sensitivity cardiac troponin I assay: A prospective study. Clinica Chimica Acta. 2025;565:119996. doi:10.1016/j.cca.2024.119996
BibTeX
@article{Desaegher2025,
author = {Desaegher, Anthony and Marin, Victor and Beauvieux, Marie-Christine and Colombiès, Brigitte and Lauga, Margaux and Alloug, Sonia and Kalkan, Selen and Castaing-Mouhica, Gladys and Lacape, Geneviève and Rucheton, Benoit and Doublet, Julien and Dabernat, Sandrine and Bats, Marie-Lise},
title = {Exploring strategies to rapidly identify false positives in high-sensitivity cardiac troponin I assay: A prospective study},
journal = {Clinica Chimica Acta},
year = {2025},
volume = {565},
pages = {119996},
doi = {10.1016/j.cca.2024.119996}
}
EUVIMED