Validation of a 0/1 h Algorithm for Rapid Diagnosis of Myocardial Infarction Using a High-Sensitivity Troponin I Assay
Clinical Chemistry · 2023
DOI: 10.1093/clinchem/hvad019 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry, 2023. Verfasst von Jonas Lehmacher und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, University Heart and Vascular Center Hamburg, University Medical Center Hamburg-Eppendorf, Hamburg, Germany. Hauptthemen laut MEDLINE: Troponin I, Myocardial Infarction. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 33 Literaturangaben. Gefördert von: German Center for Cardiovascular Research, German Ministry of Education and Research, Ernst und Berta Grimmke Stiftung.
Abstract (Originalsprache)
Background
Current guidelines recommend 0/1 h algorithms using high-sensitivity cardiac troponin (hs-cTn) for fast diagnosis of myocardial infarction (MI). Yet, for some assays, existing data is limited. We aimed to evaluate the diagnostic performance and the prognostic value of a rapid 0/1 h algorithm for the Access hs-cTnI assay.
Methods
In consecutive patients presenting with suspected MI, we measured concentrations of Access hs-cTnI at presentation and after 1 hour. Final diagnosis was adjudicated independently by 2 cardiologists. Parameters for diagnostic performance were calculated, applying the recently derived European Society of Cardiology (ESC) 0/1 h algorithm for Access hs-cTnI. Additionally, we assessed the prognostic utility of Access hs-cTnI for the composite end point of all-cause mortality and incident MI at 3 years.
Results
In 1879 patients, 257 non-ST-elevation MIs occurred. Application of the 0/1 h algorithm classified 44.5% as rule-out, 20.3% as rule-in, and triaged 35.1% to the observe group. High rule-out safety was confirmed with a sensitivity of 97.7% (95% CI, 95.0%–99.1%) and a negative predictive value of 99.3% (95% CI, 98.4%–99.7%). Rule-in capacity was moderate with a specificity of 88.0% (95% CI, 86.3%–89.6%) and a positive predictive value of 50.8% (95% CI, 45.7%–55.9%). After exclusion of patients with ST-elevation MI the results showed strong prognostic value, even after adjustment for cardiovascular risk factors and comorbidities, with adjusted hazard ratios of 2.51 (95% CI, 1.56–4.04) in the observe and 3.55 (95% CI, 2.18–5.79) in the rule-in group for the composite end point of all-cause mortality and incident MI at 3 years, compared to ruled-out patients.
Conclusion
The ESC 0/1 h algorithm for Access hs-cTnI allows safe and efficient triage of patients with suspected MI and has strong prognostic utility up to 3 years after the initial evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Cardiology, University Heart and Vascular Center Hamburg, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
- University Center of Cardiovascular Science, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
- Deutsches Zentrum für Herz-Kreislauf-Forschung (DZHK), Standort Hamburg Kiel/Lübeck, Hamburg, Germany.
- Department of Cardiology, University Heart Center Freiburg, Bad Krozingen, Germany.
- Population Health Research Department, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
- Department of Epidemiology and Preventive Medicine, School of Public Health and Preventive Medicine, Monash University, Melbourne, Australia.
ORCID-Profile: Jonas Lehmacher, Betül Toprak, Alina Goßling, Paul Michael Haller, Raphael Twerenbold, Dirk Westermann
Förderung
- German Center for Cardiovascular Research
- German Ministry of Education and Research
- Ernst und Berta Grimmke Stiftung
- German Heart Foundation
- Swiss National Science Foundation
- Swiss Heart Foundation
- Swiss Society of Cardiology
- Cardiovascular Research Foundation Basel, the University of Basel, and the University Hospital Basel
Zitieren
APA
Lehmacher, J., Toprak, B., Sörensen, N. A., Bei der Kellen, R., Goßling, A., Hartikainen, T. S., Haller, P. M., Schock, A., Twerenbold, R., Zeller, T., Blankenberg, S., Westermann, D., & Neumann, J. T. (2023). Validation of a 0/1 h Algorithm for Rapid Diagnosis of Myocardial Infarction Using a High-Sensitivity Troponin I Assay. Clinical Chemistry. https://doi.org/10.1093/clinchem/hvad019
Vancouver
Lehmacher J, Toprak B, Sörensen NA, Bei der Kellen R, Goßling A, Hartikainen TS, et al. Validation of a 0/1 h Algorithm for Rapid Diagnosis of Myocardial Infarction Using a High-Sensitivity Troponin I Assay. Clinical Chemistry. 2023. doi:10.1093/clinchem/hvad019
BibTeX
@article{Lehmacher2023,
author = {Lehmacher, Jonas and Toprak, Betül and Sörensen, Nils Arne and Bei der Kellen, Ramona and Goßling, Alina and Hartikainen, Tau Sarra and Haller, Paul Michael and Schock, Alina and Twerenbold, Raphael and Zeller, Tanja and Blankenberg, Stefan and Westermann, Dirk and Neumann, Johannes Tobias},
title = {Validation of a 0/1 h Algorithm for Rapid Diagnosis of Myocardial Infarction Using a High-Sensitivity Troponin I Assay},
journal = {Clinical Chemistry},
year = {2023},
doi = {10.1093/clinchem/hvad019}
}
EUVIMED