Validation of a 0/2‐Hour Point‐Of‐Care High Sensitivity Troponin Algorithm: A Multisite US Study
Academic Emergency Medicine · 2026-07 · 33 (7)
DOI: 10.1111/acem.70379 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Academic Emergency Medicine, online erschienen am 18.07.2026. Verfasst von Simon A Mahler und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA. Hauptthemen laut MEDLINE: Algorithms, Troponin I, Point-of-Care Systems, Myocardial Infarction. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: AHRQ HHS, NHLBI NIH HHS, NCATS NIH HHS.
Abstract (Originalsprache)
Objectives
A 0/2‐h algorithm for the i‐STAT point‐of‐care (POC) high sensitivity troponin I (hs‐cTnI) assay was recently derived in Australia. The objective of this study was to validate and optimize the performance of the 0/2‐h algorithm in a multisite U.S. Emergency Department (ED) cohort.
Methods
A prospective cohort study was conducted at three U.S. EDs (February–September 2025). Adults without STEMI and at least one hs‐cTnI ordered were accrued. Blood samples were collected simultaneously for POC hs‐cTnI measurement on an i‐STAT 1 analyzer (Abbott Laboratories) and central laboratory hs‐cTnI measurement (Beckman Coulter). The primary outcome was index myocardial infarction (MI), adjudicated by experts using clinical hs‐cTnI measures. Diagnostic performance of the Australian 0/2‐h algorithm was assessed by calculating negative and positive predictive values (NPV, PPV) with associated 95% confidence intervals. Efficacy, defined as the proportion of patients classified into the rule‐out zone, was calculated. Algorithm optimization tested modified cut points to increase efficacy while maintaining NPV ≥ 99% and achieving PPV ≥ 65%.
Results
During the study period, 578 patients with complete 0/2‐h algorithm assessments were accrued. These patients were 48% (279/578) female, 40% (233/578) non‐White, with a median age of 60 years (IQR: 50–70). Index MI occurred in 7.4% (43/578). Algorithm efficacy was 54.8% (317/578) and 8.8% (51/578) were classified to the rule‐in zone. Among patients ruled‐out, the NPV was 99.4% (95% CI: 97.7%–99.9%). The rule‐in zone was associated with a PPV of 62.7% (95% CI: 48.1%–75.9%). Among 36.3% (210/578) patients classified to the observation zone, 4.3% (9/210) had an adjudicated index MI. An optimized 0/2‐h algorithm increased efficacy to 60.0% (347/578) while achieving an NPV of 99.4% (95% CI: 97.9%–99.9%) and PPV of 74.4% (95% CI: 58.8, 86.5%) for index MI.
Conclusions
The original Australian and optimized 0/2‐h algorithms for i‐STAT POC hs‐cTnI measurement had high NPV and efficacy in a multisite U.S. cohort. Trial Registration Trial Registration: NCT06899776
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Emergency Medicine, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Department of Implementation Science, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Department of Epidemiology and Prevention, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Department of Biostatistics and Data Science, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Department of Emergency Medicine, Carolinas Medical Center, Charlotte, North Carolina, USA.
ORCID-Profile: Nicklaus P Ashburn
Förderung
- AHRQ HHS
- NHLBI NIH HHS
- NCATS NIH HHS
Zitieren
APA
Mahler, S. A., Ashburn, N. P., Villenthi, A., Pearson, D. A., Pugh, H. D., Hutchison, B. T., Supples, M. W., & Snavely, A. C. (2026). Validation of a 0/2‐Hour Point‐Of‐Care High Sensitivity Troponin Algorithm: A Multisite US Study. Academic Emergency Medicine, 33(7). https://doi.org/10.1111/acem.70379
Vancouver
Mahler SA, Ashburn NP, Villenthi A, Pearson DA, Pugh HD, Hutchison BT, et al. Validation of a 0/2‐Hour Point‐Of‐Care High Sensitivity Troponin Algorithm: A Multisite US Study. Academic Emergency Medicine. 2026;33(7). doi:10.1111/acem.70379
BibTeX
@article{Mahler2026,
author = {Mahler, Simon A and Ashburn, Nicklaus P and Villenthi, Akalya and Pearson, David A and Pugh, Hannah D and Hutchison, Benjamin T and Supples, Michael W and Snavely, Anna C},
title = {Validation of a 0/2‐Hour Point‐Of‐Care High Sensitivity Troponin Algorithm: A Multisite US Study},
journal = {Academic Emergency Medicine},
year = {2026},
volume = {33},
number = {7},
doi = {10.1111/acem.70379}
}
EUVIMED