EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Efficient Monte Carlo modeling for Gamma Knife Perfexion

Díaz Nataly, Gutiérrez Jorge, Martínez Julio, Udías José Manuel, Ibáñez Paula

Applied Radiation and Isotopes · 2026-09 · 235 · 112686

DOI: 10.1016/j.apradiso.2026.112686 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Radiation and Isotopes, 2026-09. Verfasst von Díaz Nataly und 4 weiteren Autor:innen. Beteiligte Einrichtung: Nuclear Physics Group and IPARCOS, Department of Structure of Matter, Thermal Physics and Electronics, CEI Moncloa, Universidad Complutense de Madrid…. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiosurgery. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Community of Madrid, European Commission, Universidad Complutense de Madrid.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Monte Carlo (MC) simulations constitute the most accurate tool for dosimetric analysis in small radiation fields, such as those generated by the Leksell Gamma Knife Perfexion (LGK-PFX) system. However, implementing a fully detailed model of the system is computationally demanding, both in terms of processing time and geometric complexity, due to the explicit inclusion of the 576 collimators distributed throughout the device. To address this limitation, we present an efficient and accurate model based on a single phase-space file (PSF) generated from a source-collimator simulation for each collimator size (4, 8, and 16 mm), computed using PenEasy Monte. The resulting PSF is subsequently rotated to the required orientations to reproduce the system response. This approach drastically reduces the effective computational burden, as each reference PSF is generated only once and can then be reused to simulate any desired LGK-PFX configuration. Compared with full-system models that require hundreds of hours of computation and storage on the order of terabytes, the proposed rotating-PSF strategy enables Monte Carlo simulations with moderate computational resources, making MC-based verification feasible in a clinical context. The model was validated by comparing MC-generated dose distributions with measurements performed using EBT4 radiochromic films at Ruber International Hospital (Madrid), as well as with data provided by Elekta and other MC-based studies. A 7%-0.5 mm gamma analysis of the dose profiles showed pass rates above 95%. Output factors (OFs) for the 4 mm and 8 mm collimators were 0.848 ± 0.011 and 0.889 ± 0.011, respectively, in a cubic water phantom. Overall, the rotating PSF approach preserves dosimetric accuracy while substantially improving computational efficiency, providing a practical and clinically relevant solution for dosimetry with the LGK-PFX system.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodRadiosurgeryComputer SimulationRadiometryRadiotherapy DosageRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

Gamma KnifeMonte CarloOutput factorsPerfexionSmall field dosimetry

Autor:innen und Institutionen

  • Nuclear Physics Group and IPARCOS, Department of Structure of Matter, Thermal Physics and Electronics, CEI Moncloa, Universidad Complutense de Madrid, Madrid, 28015, Spain. Electronic address: natadi02@ucm.es.
  • Hospital Ruber International, Madrid, 28034, Spain. Electronic address: jorge.gutierrez@ruberinternacional.es.
  • Hospital Ruber International, Madrid, 28034, Spain. Electronic address: juliol.martinez@ruberinternacional.es.
  • Nuclear Physics Group and IPARCOS, Department of Structure of Matter, Thermal Physics and Electronics, CEI Moncloa, Universidad Complutense de Madrid, Madrid, 28015, Spain; Instituto de Investigacion Sanitaria del Hospi…

Förderung

  • Community of Madrid
  • European Commission
  • Universidad Complutense de Madrid

Zitieren

APA

Nataly, D., Jorge, G., Julio, M., José Manuel, U., & Paula, I. (2026). Efficient Monte Carlo modeling for Gamma Knife Perfexion. Applied Radiation and Isotopes, 235, 112686. https://doi.org/10.1016/j.apradiso.2026.112686

Vancouver

Nataly D, Jorge G, Julio M, José Manuel U, Paula I. Efficient Monte Carlo modeling for Gamma Knife Perfexion. Applied Radiation and Isotopes. 2026;235:112686. doi:10.1016/j.apradiso.2026.112686

BibTeX
@article{Nataly2026,
  author = {Nataly, Díaz and Jorge, Gutiérrez and Julio, Martínez and José Manuel, Udías and Paula, Ibáñez},
  title = {Efficient Monte Carlo modeling for Gamma Knife Perfexion},
  journal = {Applied Radiation and Isotopes},
  year = {2026},
  volume = {235},
  pages = {112686},
  doi = {10.1016/j.apradiso.2026.112686}
}
Kostenlos starten →