Fachartikel in der Zeitschrift The Journal of the American Dental Association, 2023. Verfasst von Francisco Antonio Uchoa-Junior und 4 weiteren Autor:innen. Hauptthemen laut MEDLINE: Diamond, Sterilization. Die Arbeit nennt 44 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Rotary cutting instruments (RCIs) are sterilized routinely. The authors aimed to analyze the structural integrity, presence of dirt, and microbial contamination of RCIs used in clinical practice after processing. METHODS: Eighty-four RCIs (42 carbide burs, 42 diamond burs) were divided into baseline, control, and test groups. The RCIs were evaluated by means of scanning electron microscopy and microbiological analysis. Evaluation criteria included presence of structural damage, dirt, biofilm, and isolated cells and their phenotypic profile. RESULTS: The carbide burs from all groups and diamond burs from the test groups had structural damage. Dirt was observed in the baseline and test groups. Three bacterial species were isolated from 4 RCIs (9.52%). An isolated cell was observed from 1 carbide bur. Biofilm was observed on 3 RCIs (7.14%). CONCLUSIONS: RCIs should not be subjected to multiple uses; after the first clinical use they accumulate structural damage and dirt that hampers the cleaning step, causing failure in the sterilization process. PRACTICAL IMPLICATIONS: The presence of microorganisms and structural damage on the RCIs confirmed that they are not amenable to processing, a fact that characterizes them as a single-use health care product.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

DiamondSterilizationHumansSurface PropertiesMicroscopy, Electron, ScanningDental InstrumentsBiofilms

Stichwörter der Autor:innen

Dental instrumentsbiofilmmicrobiological analysisscanning electron microscopysterilization

Substanzen

Diamond

Förderung

  • Coordenação de Aperfeiçoamento de Pessoal de Nível Superior

Zitieren

APA

Uchoa-Junior, F. A., Barata, T. J. E., Leão-Vasconcelos, L. S. N. O., Ribeiro, E. L., & Tipple, A. F. V. (2023). Biofilm on and structural damage of rotary cutting instruments after 5 cycles of clinical use and processing. The Journal of the American Dental Association. https://doi.org/10.1016/j.adaj.2023.03.005

Vancouver

Uchoa-Junior FA, Barata TJE, Leão-Vasconcelos LSNO, Ribeiro EL, Tipple AFV. Biofilm on and structural damage of rotary cutting instruments after 5 cycles of clinical use and processing. The Journal of the American Dental Association. 2023. doi:10.1016/j.adaj.2023.03.005

BibTeX
@article{UchoaJunior2023,
  author = {Uchoa-Junior, Francisco Antonio and Barata, Terezinha de Jesus Esteves and Leão-Vasconcelos, Lara Stefânia Netto de Oliveira and Ribeiro, Evandro Leão and Tipple, Anaclara Ferreira Veiga},
  title = {Biofilm on and structural damage of rotary cutting instruments after 5 cycles of clinical use and processing},
  journal = {The Journal of the American Dental Association},
  year = {2023},
  doi = {10.1016/j.adaj.2023.03.005}
}
Kostenlos starten →