Effects of Repeated Use and Sterilization on the Wear of Zirconia Implant Drills: A SEM‐Based Analysis

Vasilios Alevizakos iD, Richard Mosch, Ann-Christin Platte, Constantin von See

Clinical and Experimental Dental Research · 2025

DOI: 10.1002/cre2.70088 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical and Experimental Dental Research, 2025. Verfasst von Vasilios Alevizakos und 3 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Digital Technologies in Dentistry and CAD/CAM, Danube Private University, Krems an der Donau, Austria. Hauptthemen laut MEDLINE: Zirconium, Sterilization, Dental Implants, Dental Instruments. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study evaluated the effects of repeated use and sterilization on the wear and cutting‐edge integrity of zirconia implant drills. Materials and Methods Sixty zirconium dioxide drills (Z‐Systems AG) with diameters of 2.3, 3.75, and 4.25 mm were tested. Drilling was performed in porcine mandibular bone under standardized conditions: 800 rpm, 50 Ncm torque, and 20 N axial pressure. Drills were divided into two groups: Group 1 (sterilized but unused) and Group 2 (30 drilling cycles with reprocessing). Wear was assessed using scanning electron microscopy (SEM) at 1000x magnification, applying a three‐grade scoring system. Statistical analysis was performed using the Mann–Whitney U test ( p < 0.05).

Results

Drills subjected to 30 cycles showed significantly higher wear grades (Grade 2–3) compared to unused drills (Grade 0–1) ( p < 0.001). The mean wear grades increased from 0.3 to 2.6 for 2.3 mm, from 0.4 to 2.7 for 3.75 mm, and from 0.2 to 2.7 for 4.25 mm drills. Effect sizes ( r = 0.88–0.90) confirmed a strong relationship between repeated use and wear.

Conclusions

Zirconia drills show significant wear after 30 cycles. Although they offer potential as an alternative to steel drills, further research is needed to optimize cost‐effectiveness and clinical durability. Clinical Trial Registration Not applicable.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumSterilizationDental ImplantsDental InstrumentsAnimalsMicroscopy, Electron, ScanningSwineMaterials TestingSurface PropertiesEquipment Reuse

Stichwörter der Autor:innen

dental implantdrill wearsterilizationzirconium oxide

Substanzen

Zirconiumzirconium oxideDental Implants

Autor:innen und Institutionen

  • Research Center for Digital Technologies in Dentistry and CAD/CAM, Danube Private University, Krems an der Donau, Austria.

ORCID-Profile: Vasilios Alevizakos

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Alevizakos, V., Mosch, R., Platte, A. C., & von See, C. (2025). Effects of Repeated Use and Sterilization on the Wear of Zirconia Implant Drills: A SEM‐Based Analysis. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70088

Vancouver

Alevizakos V, Mosch R, Platte AC, von See C. Effects of Repeated Use and Sterilization on the Wear of Zirconia Implant Drills: A SEM‐Based Analysis. Clinical and Experimental Dental Research. 2025. doi:10.1002/cre2.70088

BibTeX
@article{Alevizakos2025,
  author = {Alevizakos, Vasilios and Mosch, Richard and Platte, Ann-Christin and von See, Constantin},
  title = {Effects of Repeated Use and Sterilization on the Wear of Zirconia Implant Drills: A SEM‐Based Analysis},
  journal = {Clinical and Experimental Dental Research},
  year = {2025},
  doi = {10.1002/cre2.70088}
}
Kostenlos starten →