Clinical applications with expired EBT-XD radiochromic film in radiotherapy

O A García-Garduño, Y Ávila-Muñoz, I Gamboa-deBuen, J M Lárraga-Gutiérrez

Physical and Engineering Sciences in Medicine · 2026

DOI: 10.1007/s13246-026-01773-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2026. Verfasst von O A García-Garduño und 3 weiteren Autor:innen. Beteiligte Einrichtung: Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, México City, 14269, Mexico. oagarciag@innn.edu.mx. Hauptthemen laut MEDLINE: Film Dosimetry. Die Arbeit nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Radiochromic films are widely used in radiotherapy dosimetry for their high spatial resolution and near-water-equivalence. Nonetheless, these films are frequently purchased in bulk, leading clinical departments to accumulate stock beyond the manufacturer’s expiration date. Assessing the dosimetric performance of expired films may confer economic and environmental advantages. This work evaluated whether expired EBT-XD radiochromic films remain suitable for specific radiotherapy dosimetry applications. Two expired batches were compared with a non-expired batch. Dose–response curves were generated using a 6 MV photon beam from a clinical linear accelerator. The films were tested for small-field output factor measurements and patient-specific quality assurance (PSQA) in volumetric-modulated arc therapy (VMAT) and stereotactic radiosurgery (SRS), employing gamma-index analysis. Results indicated that expired films exhibit reduced sensitivity relative to unexpired films, particularly at lower dose levels. However, acceptable dosimetric agreement with unexpired films was observed for small-field output factors and high-dose SRS PSQA cases, provided appropriate recalibration was applied. More pronounced discrepancies emerged in low-dose VMAT PSQA measurements, where diminished film sensitivity may have influenced the film’s response. These findings suggest that expired EBT-XD radiochromic films may still provide useful dosimetric information as relative dosimeters for selected radiotherapy measurements, especially when measurements fall within the film’s recommended dose range, and films are properly stored and recalibrated.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Film DosimetryHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiosurgery

Stichwörter der Autor:innen

Expired radiochromic filmFilm dosimetrySmall-field dosimetryUncertaintyVMAT

Autor:innen und Institutionen

  • Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, México City, 14269, Mexico. oagarciag@innn.edu.mx.
  • Facultad de Ciencias, Universidad Nacional Autónoma de México, Ciudad Universitaria, México City, Mexico.
  • Instituto de Ciencias Nucleares, Universidad Nacional Autónoma de México, Av. Universidad 3000, México City, 04510, Mexico.
  • Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, México City, 14269, Mexico. jlarraga@innn.edu.mx.

Zitieren

APA

García-Garduño, O. A., Ávila-Muñoz, Y., Gamboa-deBuen, I., & Lárraga-Gutiérrez, J. M. (2026). Clinical applications with expired EBT-XD radiochromic film in radiotherapy. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-026-01773-9

Vancouver

García-Garduño OA, Ávila-Muñoz Y, Gamboa-deBuen I, Lárraga-Gutiérrez JM. Clinical applications with expired EBT-XD radiochromic film in radiotherapy. Physical and Engineering Sciences in Medicine. 2026. doi:10.1007/s13246-026-01773-9

BibTeX
@article{GarcaGarduo2026,
  author = {García-Garduño, O A and Ávila-Muñoz, Y and Gamboa-deBuen, I and Lárraga-Gutiérrez, J M},
  title = {Clinical applications with expired EBT-XD radiochromic film in radiotherapy},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2026},
  doi = {10.1007/s13246-026-01773-9}
}
Kostenlos starten →